Aim: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters.
Methods: Method comparison (according to CLSI EP09-A3), carryover (according to CLSI H57-A), APTT sensitivity to heparin (according to CLSI H47-A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15-A3) and accuracy. The assays tested on sthemO 301 were: PT, APTT (silica and kaolin activators), fibrinogen (Fib), thrombin time (TT), chromogenic and clotting protein C (PC) activity, and von Willebrand factor antigen (VWF:Ag) levels.
Results: All intra-assay and inter-assay precision CVs were below the maximal precision limit proposed by the French Group for Hemostasis and Thrombosis (GFHT). Accuracy was verified with bias below GFHT criteria and most Z-scores were between -2 and +2. No clinically relevant carryover was detected. Silica APTT reagent sensitivity to unfractionated heparin was moderate, as expected. Productivity results were consistent over the 10 repeats performed. The overall agreement between the two systems was excellent for all assays, with Spearman rank correlation coefficient all above 0.9 and slopes of Passing-Bablok correlation near 1 and intercepts close to 0.
Conclusion: For the methods tested, sthemO 301 system met all the criteria to implement a novel coagulation analyzer in the laboratory and result comparability with STA R Max® 2 was good.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10431393 | PMC |
http://dx.doi.org/10.1002/jcla.24929 | DOI Listing |
Haemophilia
July 2025
Laboratoire De Biologie Médicale (Secteur Hémostase), C.H. De Versailles André Mignot, Le Chesnay, France.
Blood Coagul Fibrinolysis
June 2025
Section of Clinical Biochemistry, University of Verona, Verona, Italy.
Diagnosis (Berl)
August 2025
Section of Clinical Biochemistry, 19051 University of Verona, Verona, Italy.
Objectives: We performed a comparative evaluation of analytical performance between the novel bench-top Stago sthemO 301 hemostasis analyzer and the Werfen ACL TOP 750 coagulometer using routine plasma samples.
Methods: A minimum of 100 fresh plasma samples per test were analyzed using both analytical systems. The tests included prothrombin time (PT) expressed as international normalized ratio (INR) and seconds (s), activated partial thromboplastin time (APTT) in ratio and seconds, fibrinogen, antithrombin and D-dimer.
J Clin Lab Anal
June 2023
Laboratoire d'Hématologie-Hémostase, Centre Hospitalo-Universitaire de Rennes Pontchaillou, Rennes, France.
Aim: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters.
Methods: Method comparison (according to CLSI EP09-A3), carryover (according to CLSI H57-A), APTT sensitivity to heparin (according to CLSI H47-A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15-A3) and accuracy.