Article Abstract

Background And Purpose: Efficient workflows for adaptive proton therapy are of high importance. This study evaluated the possibility to replace repeat-CTs (reCTs) with synthetic CTs (sCTs), created based on cone-beam CTs (CBCTs), for flagging the need of plan adaptations in intensity-modulated proton therapy (IMPT) treatment of lung cancer patients.

Materials And Methods: Forty-two IMPT patients were retrospectively included. For each patient, one CBCT and a same-day reCT were included. Two commercial sCT methods were applied; one based on CBCT number correction (Cor-sCT), and one based on deformable image registration (DIR-sCT). The clinical reCT workflow (deformable contour propagation and robust dose re-computation) was performed on the reCT as well as the two sCTs. The deformed target contours on the reCT/sCTs were checked by radiation oncologists and edited if needed. A dose-volume-histogram triggered plan adaptation method was compared between the reCT and the sCTs; patients needing a plan adaptation on the reCT but not on the sCT were denoted false negatives. As secondary evaluation, dose-volume-histogram comparison and gamma analysis (2%/2mm) were performed between the reCT and sCTs.

Results: There were five false negatives, two for Cor-sCT and three for DIR-sCT. However, three of these were only minor, and one was caused by tumour position differences between the reCT and CBCT and not by sCT quality issues. An average gamma pass rate of 93% was obtained for both sCT methods.

Conclusion: Both sCT methods were judged to be of clinical quality and valuable for reducing the amount of reCT acquisitions.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10314284PMC
http://dx.doi.org/10.1016/j.phro.2023.100459DOI Listing

Publication Analysis

Top Keywords

proton therapy
12
adaptive proton
8
lung cancer
8
rect
8
sct methods
8
performed rect
8
plan adaptation
8
false negatives
8
sct
5
clinical evaluation
4

Similar Publications

Carbon ion combined photon radiotherapy induces ferroptosis via NCOA4-mediated ferritinophagy in glioblastoma.

Redox Biol

September 2025

Department of Radiation Oncology, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai, 201321, China; Shanghai Key Laboratory of Radiation Oncology, Shanghai, 201321, China; Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai 201321,

Glioblastoma (GBM), the most prevalent and lethal primary malignancy of the central nervous system, remains refractory to conventional photon radiotherapy due to inherent limitations in dose distribution. Although carbon ion radiotherapy offers distinct advantages, including its characteristic Bragg peak deposition and superior relative biological effectiveness, its clinical application is constrained by high costs and increased toxicity. This study explores the radiobiological interactions underlying a mixed carbon ion-photon irradiation regimen, a promising strategy in advanced particle therapy.

View Article and Find Full Text PDF

Background: The most common cause of acid-peptic diseases (APDs) is Helicobacter Pylori (H. pylori) infection. Conventionally, proton-pump inhibitors (PPIs) are used to manage hyperacidity and dyspepsia.

View Article and Find Full Text PDF

Purpose: Cranial irradiation is associated with health-related quality of life (HRQoL) deficits in childhood cancer survivors. We investigated the relationship between radiation dose to brain substructures and HRQoL in children with brain tumors treated with proton beam therapy (PBT).

Methods: Data were obtained from children in the Pediatric Proton/Photon Consortium Registry who received PBT for primary brain tumors between 2015 and 2021.

View Article and Find Full Text PDF