Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Background: Image optimization is a key step in clinical nuclear medicine, and phantoms play an essential role in this process. However, most phantoms do not accurately reflect the complexity of human anatomy, and this presents a particular challenge when imaging endocrine glands to detect small (often subcentimeter) tumors. To address this, we developed a novel phantom for optimization of positron emission tomography (PET) imaging of the human pituitary gland. Using radioactive 3D printing, phantoms were created which mimicked the distribution of C-methionine in normal pituitary tissue and in a small tumor embedded in the gland (i.e., with no inactive boundary, thereby reproducing the in vivo situation). In addition, an anatomical phantom, replicating key surrounding structures [based on computed tomography (CT) images from an actual patient], was created using material extrusion 3D printing with specialized filaments that approximated the attenuation properties of bone and soft tissue.
Results: The phantom enabled us to replicate pituitary glands harboring tumors of varying sizes (2, 4 and 6 mm diameters) and differing radioactive concentrations (2 ×, 5 × and 8 × the normal gland). The anatomical phantom successfully approximated the attenuation properties of surrounding bone and soft tissue. Two iterative reconstruction algorithms [ordered subset expectation maximization (OSEM); Bayesian penalized likelihood (BPL)] with a range of reconstruction parameters (e.g., 3, 5, 7 and 9 OSEM iterations with 24 subsets; BPL regularization parameter (β) from 50 to 1000) were tested. Images were analyzed quantitatively and qualitatively by eight expert readers. Quantitatively, signal was the highest using BPL with β = 50; noise was the lowest using BPL with β = 1000; contrast was the highest using BPL with β = 100. The qualitative review found that accuracy and confidence were the highest when using BPL with β = 400.
Conclusions: The development of a bespoke phantom has allowed the identification of optimal parameters for molecular pituitary imaging: BPL reconstruction with TOF, PSF correction and a β value of 400; in addition, for small (< 4 mm) tumors with low contrast (2:1 or 5:1), sensitivity may be improved using a β value of 100. Together, these findings should increase tumor detection and confidence in reporting scans.