A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Evaluation of a smartphone-based colorimetric method for urinalysis dipstick readings in cats. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objectives: The aim of the study was to compare the diagnostic performances of a smartphone-based colorimetric method (SBCM) for urinalysis with a semi-automated point-of-care (POC) analyser using standardised solutions and cat urine.

Methods: Artificial solutions (negative and positive quality controls, and purposely designed artificial urine) and natural urine from 216 cats were used. Two urine reagent strips were simultaneously dipped in each sample. One dipstick was read by the SBCM and the other by the POC analyser at the same time. Results for pH, proteins, bilirubin, 'blood', glucose and ketones were considered. Overall agreement and sensitivity, specificity and accuracy of the SBCM were determined based on selected cut-offs.

Results: For the artificial solutions, 80 comparisons were obtained for each analyte and each expected concentration. The overall agreement (exactly the same result) between the two methods was 78.4%. SBCM sensitivity, specificity and accuracy were 99.0%, 100% and 99.3%, respectively. The correlation between the two methods was almost perfect (Cohen's kappa coefficient = 0.9851). For natural urine samples, the overall agreement (including pH) was 68.6%. Using optimal cut-offs for the SBCM determined from the results of analysis of artificial solutions, the sensitivity, specificity and accuracy of the SBCM were 100%, 76.02% and 80.5%, respectively. In this situation, the correlation between the two methods was moderate (Cohen's kappa coefficient = 0.5401). This was mostly due to a high rate of false-positive results for bilirubin (61.1%).

Conclusions And Relevance: With proper cut-off use (ie, considering positive or negative results) the SBCM evaluated here has a perfect sensitivity and appropriate diagnostic performances for proteins, 'blood', glucose and ketones. Based on these experimental data, this method appears suitable for dipstick urinalysis but positive results for bilirubin and proteins have to be confirmed.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10811987PMC
http://dx.doi.org/10.1177/1098612X231171434DOI Listing

Publication Analysis

Top Keywords

artificial solutions
12
sensitivity specificity
12
specificity accuracy
12
smartphone-based colorimetric
8
colorimetric method
8
diagnostic performances
8
poc analyser
8
natural urine
8
'blood' glucose
8
glucose ketones
8

Similar Publications