Article Abstract

 Due to the potential risk of COVID-19 transmission during endonasal surgery, we studied methods to reduce droplet and aerosol generation during these procedures.  Droplet spread was assessed using ultraviolet light and a camera that detected fluorescence in the operative field and surgeon's personal protective equipment. Density of aerosols sized <10 µm was measured using a photometric particle counter. We designed a face-mounted negative-pressure mask placed on the patients' face during endoscopic endonasal surgery. Sixteen patients were recruited between October 2020 and March 2021 and randomly assigned to the mask and no-mask groups. We compared droplet spread and aerosols generated in both groups, with copious irrigation and continuous suction during drilling forming the mainstay of surgical technique in all cases.  Droplet contamination due to direct spillage of fluorescein from the syringe was noted in two patients. Aerosol density rose during sphenoid drilling in both groups, with no significant difference when continuous suction and irrigation were employed (1.27 times vs. 1.07 times the baseline,  = 0 ). Aerosol density rose significantly when suction and irrigation were interrupted in the no-mask group (44.9 times vs. 1.2 times,  = 0 , which was not seen when the mask was used.  Aerosol generation increases during drilling in endonasal procedures and is a concern during this pandemic. The use of a rigid suction close to the drill along with copious irrigation is effective in reducing aerosol spread. The use of a negative pressure mask provides additional safety when inadvertent blockage of suction and inadequate irrigation occur.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10171932PMC
http://dx.doi.org/10.1055/a-1774-6091DOI Listing

Publication Analysis

Top Keywords

negative-pressure face-mounted
4
face-mounted system
4
system reduces
4
reduces aerosol
4
aerosol spread
4
spread endonasal
4
endonasal endoscopic
4
endoscopic surgery
4
surgery  due
4
 due potential
4

Similar Publications

 Due to the potential risk of COVID-19 transmission during endonasal surgery, we studied methods to reduce droplet and aerosol generation during these procedures.  Droplet spread was assessed using ultraviolet light and a camera that detected fluorescence in the operative field and surgeon's personal protective equipment. Density of aerosols sized <10 µm was measured using a photometric particle counter.

View Article and Find Full Text PDF

Background: The COVID-19 pandemic has revealed deficiencies in the adequacy of personal protective equipment (PPE) for healthcare workers. Endoscopic endonasal skull base surgery is thought to be among the highest-risk aerosol-generating procedures for surgeons and operating room personnel.

Objective: To validate the efficacy and clinical feasibility of a novel surgical device.

View Article and Find Full Text PDF

Objective: The authors developed a negative-pressure, patient face-mounted antechamber and tested its efficacy as a tool for sequestering aerated particles and improving the safety of endonasal surgical procedures.

Methods: Antechamber prototyping was performed with 3D printing and silicone-elastomer molding. The lowest vacuum settings needed to meet specifications for class I biosafety cabinets (flow rate ≥ 0.

View Article and Find Full Text PDF