Article Abstract

Introduction: Portable spirometers are commonly used in longitudinal epidemiological studies to measure and track the forced expiratory volume in first second (FEV1) and forced vital capacity (FVC). During the course of the study, it may be necessary to replace spirometers with a different model. This raise questions regarding the comparability of measurements from different devices. We examined the correlation, mean differences and agreement between two different spirometers, across diverse populations and different participant characteristics.

Methods: From June 2015 to Jan 2018, a total of 4,603 adults were enrolled from 628 communities in 18 countries and 7 regions of the world. Each participant performed concurrent measurements from the MicroGP and EasyOne spirometer. Measurements were compared by the intra-class correlation coefficient (ICC) and Bland-Altman method.

Results: Approximately 65% of the participants achieved clinically acceptable quality measurements. Overall correlations between paired FEV1 (ICC 0.88 [95% CI 0.87, 0.88]) and FVC (ICC 0.84 [0.83, 0.85]) were high. Mean differences between paired FEV1 (-0.038 L [-0.053, -0.023]) and FVC (0.033 L [0.012, 0.054]) were small. The 95% limits of agreement were wide but unbiased (FEV1 984, -1060; FVC 1460, -1394). Similar findings were observed across regions. The source of variation between spirometers was mainly at the participant level. Older age, higher body mass index, tobacco smoking and known COPD/asthma did not adversely impact on the inter-device variability. Furthermore, there were small and acceptable mean differences between paired FEV1 and FVC z-scores using the Global Lung Initiative normative values, suggesting minimal impact on lung function interpretation.

Conclusions: In this multicenter, diverse community-based cohort study, measurements from two portable spirometers provided good correlation, small and unbiased differences between measurements. These data support their interchangeable use across diverse populations to provide accurate trends in serial lung function measurements in epidemiological studies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC10021326PMC
http://dx.doi.org/10.1371/journal.pgph.0000141DOI Listing

Publication Analysis

Top Keywords

paired fev1
12
differences agreement
8
spirometers diverse
8
diverse community-based
8
portable spirometers
8
epidemiological studies
8
diverse populations
8
differences paired
8
lung function
8
measurements
7

Similar Publications

Paternal prepubertal passive smoke exposure is related to impaired lung function trajectories from childhood to middle age in their offspring.

Thorax

September 2025

Allergy and Lung Health Unit, Centre for Epidemiology and Biostatistics, Melbourne School of Population and Global Health, The University of Melbourne, Melbourne, Victoria, Australia.

Introduction: Paternal prepubertal passive smoke exposure may increase the risk of childhood asthma. However, its association with impaired lung function trajectories at risk of chronic obstructive pulmonary disease in offspring was not investigated. We assessed the association between paternal prepubertal passive smoke exposure and lung function from childhood to middle age in their offspring.

View Article and Find Full Text PDF

Background: Air pollution exposure during pregnancy and childhood has been linked to lung development deficits in children, however, few studies have identified potential windows of susceptibility to air pollution exposure from conception to early childhood on lung function.

Objectives: To identify potential windows of susceptibility to the effects of prenatal and childhood exposure to air pollution and lung function.

Methods: We included 1029 mother-child pairs from the INMA (INfancia y Medio Ambiente) birth cohort.

View Article and Find Full Text PDF

Purpose: To evaluate the feasibility of combining pulsed field ablation through irreversible electroporation (IRE) or pulsed electric field (PEF) with stereotactic body radiation therapy (SBRT) delivered within 15 days for lung tumors.

Materials And Methods: A prospective pilot study was designed for the treatment of lung tumors with pulsed field ablation followed by 12 Gy in 1 fraction. Percutaneous ablation via paired electrode probes (IRE) or a single probe system (PEF) was allowed on the trial; the intent was complete ablation.

View Article and Find Full Text PDF

Background: Inhaled corticosteroids (ICS) are widely used to treat chronic obstructive pulmonary disease (COPD), but treatment responses vary among individuals. Identifying biomarkers that can improve our understanding of disease mechanisms and help predict ICS responsiveness is urgently needed. Extracellular vesicles (EVs), key mediators of intercellular communication, may offer novel insights and serve as a potential biomarker source.

View Article and Find Full Text PDF

Background: Although web-based intervention programs are effective in changing health behavior, evidence of their effectiveness in relation to air pollution and respiratory health in children is lacking. We assessed the effects of web-based behavioral intervention on exposure to fine particulate matter (PM ≤ 2.5 μm in diameter [PM]), lung function, and airway inflammation in children.

View Article and Find Full Text PDF