Article Abstract

Background: Orthopaedic devices comprise nearly 20% of devices on the market and 12% to 20% of these devices undergo a recall within 10 years. More than 95% of these devices are approved without supporting clinical data through the Food and Drug Administration's 510(k) pathway. The risk of recall of orthopaedic arthroplasty devices approved through the 510(k) pathway has not been previously studied.

Methods: The FDA 510(k) database was queried for orthopaedic devices approved between January 01, 2008 and December 31, 2018 and subsequently codified to hip and knee arthroplasty devices using product codes. The database included 904 arthroplasty devices during the study period, with hip and knee making up 53.7% (485) and 46.3% (419) of devices, respectively. Information regarding numbers, dates, and reasons for recall were recorded. Cumulative incidence function was conducted to compare the risk of recall between hip and knee arthroplasty.

Results: In total, 94 (19.4%) hip and 85 (20.3%) knee devices were recalled. The hazard of recall by 10 years for hip and knee arthroplasty devices was approximately 24%, with no statistical differences between each region. The most common causes of recall were process control and device design, accounting for 29.6% and 26.3% of recalls, respectively, with no significant difference between study groups.

Conclusion: The risk of recall for arthroplasty devices is more than that previously understood. Improved postmarket surveillance strategies along with increased physician participation in detecting and reporting device safety issues are necessary to strengthen patient safety.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arth.2023.01.068DOI Listing

Publication Analysis

Top Keywords

arthroplasty devices
24
risk recall
16
hip knee
16
devices
13
devices approved
12
orthopaedic devices
8
20% devices
8
recall years
8
510k pathway
8
knee arthroplasty
8

Similar Publications

Bone mineral density assessment using radiofrequency echographic multispectrometry (REMS) in patients before and after total hip replacement.

Osteoporos Int

September 2025

Department of Rheumatology, First Faculty of Medicine, Charles University, Katerinska 32, Prague, 121 08, Czech Republic.

Unlabelled: REMS-BMD by radiofrequency echographic multispectrometry is primarily determined by a patient's BMI, age, and sex. Only about 2.8% of the changes in femoral neck REMS-BMD can be attributed to replacement of the total hip with metal implants.

View Article and Find Full Text PDF

Healthcare-associated infections (HAIs) continue to pose major risks to pediatric and neonatal patients, whose immature immune systems and unique vulnerabilities demand tailored infection prevention strategies. Traditional methods, including chemical disinfectants, procedural protocols, and physical hygiene measures, have contributed to reductions in HAIs but remain limited by human error, environmental toxicity, and the rise of antimicrobial resistance. Advances in disinfection technologies, particularly ultraviolet-C (UV-C) systems, offer promising new avenues for safer, more effective pathogen control.

View Article and Find Full Text PDF

Introduction: anatomical deformities such as developmental dysplasia of the hip (DDH) and Perthes disease represent a challenge for reconstruction. The use of 3D-printed models can be helpful for assessing the deformity, bone mass, implant size, and orientation.

Objectives: to prospectively evaluate the outcomes of 3D simulation in primary total hip arthroplasty.

View Article and Find Full Text PDF

Background: Accurate acetabular cup orientation in total hip arthroplasty (THA) is crucial for successful outcomes. Intraoperative fluoroscopy may be used to evaluate acetabular cup placement. This study aimed to evaluate the accuracy of purely visual estimation of cup inclination and anteversion using intraoperative fluoroscopy, considering different surgeon experience levels and cup designs.

View Article and Find Full Text PDF

Image-based robotics enhance precision and efficiency in lateral UKA: A comparative study of 135 UKAs.

Knee Surg Sports Traumatol Arthrosc

September 2025

Orthopaedics Surgery and Sports Medicine Department, FIFA Medical Center of Excellence, Croix-Rousse Hospital, Hospices Civils de Lyon, Lyon North University Hospital, Lyon, France.

Purpose: Robotic-assisted lateral unicompartmental knee arthroplasty (UKA) remains technically demanding due to the complex biomechanics of the lateral compartment. Image-based (IBRA) and imageless (ILRA) robotic systems have both demonstrated superior accuracy compared to conventional mechanical instrumentation, but have not yet been directly compared in lateral UKA. This study aimed to evaluate their respective accuracy and surgical efficiency.

View Article and Find Full Text PDF