Article Abstract

Background: Published analyses of never-irradiated, ethylene oxide (EtO)-sterilized tibial inserts and EtO- and gas plasma (GP)-sterilized acetabular ultra-high molecular weight polyethylene (UHMWPE) retrievals demonstrated minimal UHMWPE in vivo oxidation, whereas another analysis of EtO-sterilized acetabular liners found elevated oxidation linked with in vivo stresses. This study explored whether never-irradiated UHMWPE bearings are (1) oxidized by the in vivo environment, and (2) more likely to oxidize in higher-stress articulations (knee, ankle, shoulder).

Methods: An institutional review board-approved retrieval archive was queried for never-irradiated, EtO- and GP-sterilized UHMWPE bearings received at revision from 2001 to 2021. A total of 193 EtO-sterilized and 112 GP-sterilized conventional UHMWPE retrievals were analyzed (0 to 25 years in vivo; 133 hip, 144 knee, 18 ankle, and 10 shoulder). Retrieved implants were evaluated for in vivo damage and analyzed for trans-vinylene and ketone content by Fourier transform infrared spectroscopy (FTIR). Twelve never-implanted EtO-sterilized tibial knee inserts, (shelf-aged 5 to 19 years) were non-oxidized controls. Mechanical properties of 3 never-implanted and 3 retrieved tibial inserts were evaluated by ASTM Type-V uniaxial tensile testing. Statistical analyses evaluated correlations between time in vivo and oxidation, and compared mean oxidation rates by articulation.

Results: Burnishing was the most common clinical damage for all articulations. Eight retrievals exhibited oxidation-related fatigue damage. All retrievals were validated as never-irradiated (median trans-vinylene index [TVI] = 0.000). Maximum ketone oxidation in retrievals correlated with in vivo time (p < 0.001). Thirty-seven percent of retrievals exhibited UHMWPE (subsurface) oxidation, most frequently ankle, knee, and glenoid inserts. Tensile properties differed between retrieved and never-implanted inserts, changing with oxidation. The oxidation rate differed significantly among the articulations (p < 0.001).

Conclusions: This study cohort confirmed the presence of in vivo oxidation in some non-irradiation-sterilized UHMWPE bearings, with higher-stress articulations (knee, ankle, shoulder) showing evidence of oxidation more frequently and having significantly higher oxidation rates than hips. Mechanical properties degraded by oxidation led to fatigue damage in 8 retrievals after a long duration in vivo.

Clinical Relevance: Conventional EtO- or GP-sterilized UHMWPE bearings are at minimal risk for fatigue damage secondary to oxidation. However, higher stresses and longer time in vivo (more cycles of use) can lead to increased wear, oxidation, and fatigue damage.

Download full-text PDF

Source
http://dx.doi.org/10.2106/JBJS.22.00571DOI Listing

Publication Analysis

Top Keywords

uhmwpe bearings
20
fatigue damage
16
oxidation
15
vivo oxidation
12
knee ankle
12
vivo
10
uhmwpe
9
never-irradiated uhmwpe
8
eto-sterilized tibial
8
tibial inserts
8

Similar Publications

Patients undergo total knee arthroplasty (TKA) at younger ages with the expectation that the devices will perform well over two to three decades. During this time, the ultra-high molecular weight polyethylene (UHMWPE) bearing material properties of the implant may change due to aging induced by radiation and oxygen diffusion or other effects. Vitamin E or other antioxidants are promoted since several years to improve the oxidation resistance of UHMWPE.

View Article and Find Full Text PDF

Joint arthroplasty bearing materials must maintain a balance between wear resistance, toughness, and oxidation resistance. Antioxidant-doped polyethylene has been introduced to stabilize free radicals resulting from the cross-linking process while avoiding mechanical property losses associated with previous generations of highly cross-linked polyethylene. Furthermore, the antioxidant should prevent or greatly reduce oxidation occurring in vivo.

View Article and Find Full Text PDF

Cobalt-chromium-molybdenum (Co-Cr-Mo) femoral components are widely used in total knee arthroplasty (TKA) due to their mechanical strength and wear performance. However, concerns regarding corrosion, metal ion release, and hypersensitivity have prompted the development of alternative materials, including nitrided titanium-aluminum-vanadium (nTi-6Al-4V). This study aimed to compare the simulated wear performance of Co-Cr-Mo and nTi-6Al-4V femoral components when articulated against conventional polyethylene articular surface bearings and evaluate the fatigue performance of nTi-6Al-4V components.

View Article and Find Full Text PDF

Super Hydrophobic UHMWPE/PTFE/PVA Composites with Low Friction: Preparation and Wear Mechanism.

Polymers (Basel)

June 2025

Aviation Key Laboratory of Science and Technology on Generic Technology of Self-Lubricating Spherical Plain Bearing, Yanshan University, Qinhuangdao 066004, China.

This study develops novel superhydrophobic UHMWPE/PTFE/PVA composites via hot-pressing sintering to achieve ultra-low friction and enhanced wear resistance. The ternary system synergistically combines UHMWPE's mechanical stability, PTFE's lubricity, and PVA's dispersion/binding capability. Results show PTFE disrupts UHMWPE crystallization, reducing melting temperature by 2.

View Article and Find Full Text PDF

Knee arthroplasty (KA) represents a transformative milestone in the management of degenerative knee conditions, significantly improving patient mobility and quality of life. Over the decades, material innovations have driven advancements in implant design, addressing challenges such as wear, biocompatibility, and longevity. This review provides a comprehensive evaluation of traditional and cutting-edge materials used in KA, analyzing their properties, clinical outcomes, and economic implications while identifying future research directions.

View Article and Find Full Text PDF