A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Disposable laryngoscope intubation to reduce equipment failure in an emergency out of OR setting - a quality control case study. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Reusable laryngoscopes have been reported to be superior to disposable laryngoscopes with plastic blades during emergent intubations. Surprisingly, at our institution a quality reporting system revealed a high number of equipment failures with reusable laryngoscopes in an emergency out-of-OR (operating room) setting. As recent studies indicated an improved quality of disposable laryngoscopes, we hypothesized that a thoroughly evaluated disposable laryngoscope would result in less equipment failure in an emergency out-of-OR setting.

Methods: To perform a more standardized and time efficient analysis, four distinct disposable laryngoscope blade/handle configurations were trialed during standard intubations (n = 4 × 30) in the OR by experienced anesthesia providers who completed a 6-question, Likert-scale/open-ended survey for product evaluation. The 'best' disposable blade was implemented in an emergency out-of-OR setting and equipment failure rates were monitored over a 3-year period.

Results: Different disposable laryngoscopes were equal regarding sturdiness, illumination and airway visualization. The laryngoscope with the highest overall score was significantly higher scored than the laryngoscope with the lowest overall score. All disposable laryngoscopes were more cost effective than the reusable ones, and the top scored laryngoscope demonstrated the highest 5-year cost-saving ($210 K). Implementation of the top scored disposable laryngoscope into an emergency out-of-OR setting reduced the equipment failure incidence from high 20s to 0.

Conclusion: Disposable laryngoscopes are cost effective and superior to reusable laryngoscopes in an emergency out-of-OR setting. We demonstrate that the implementation of a disposable laryngoscope in the emergency out-of-OR setting resulted in a near elimination of equipment related quality submissions which ultimately enhances patient safety.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9830876PMC
http://dx.doi.org/10.1186/s12871-022-01956-3DOI Listing

Publication Analysis

Top Keywords

emergency out-of-or
24
disposable laryngoscope
20
disposable laryngoscopes
20
equipment failure
16
out-of-or setting
16
reusable laryngoscopes
12
disposable
11
failure emergency
8
laryngoscopes
8
laryngoscopes emergency
8

Similar Publications