A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Results from the First-in-Human Study of the Caterpillar™ Arterial Embolization System. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: To assess occlusion success and adverse events associated with the use of a self-expanding device for peripheral artery embolization.

Methods: This prospective, single-arm, feasibility study was conducted using the Caterpillar™ Arterial Embolization Device composed of opposing nitinol fibers and a flow-occluding membrane. Twenty patients (24 embolization sites) were treated at four investigational centers in New Zealand and Australia and followed for 30 days. Embolization sites included mesenteric, accessory renal, and iliac arteries and their branches. Primary outcome measures were peri-procedural occlusion confirmed by angiography and freedom from device-related serious adverse events (SAEs) at 30 days. Secondary observations included time to occlusion and assessment of adverse events.

Results: Peri-procedural occlusion was 100%, and freedom from a device-related SAE was 94.7% at 30 days. One patient had abdominal bloating that required hospitalization deemed possibly related to the device or procedure. Twenty-two of 24 embolization sites were occluded with one device (91.7%). Mean procedure duration was 11.7 ± 8.6 min (device deployment time: 1.8 ± 1.0 min), and mean fluoroscopy time was 241 ± 290.7 s. All embolization sites occluded during the procedure with 62.5% occluded within three minutes and 91.6% occluded within ten minutes. No devices migrated or required re-embolization. Freedom from device- and procedure-related adverse events was 84.2%. One patient died from aortic rupture during a subsequent adjunctive abdominal aortic endovascular procedure deemed unrelated to the embolization device or procedure.

Conclusions: This first-in-human study of the Caterpillar embolization device achieved peri-procedural occlusion in all patients with a 94.7% freedom from device-related SAE at 30 days.

Level Of Evidence: Level 2b-prospective, multicenter, single-arm, first-in-human clinical study. Pre-specified endpoints were analyzed using descriptive statistics.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9713177PMC
http://dx.doi.org/10.1007/s00270-022-03300-1DOI Listing

Publication Analysis

Top Keywords

adverse events
12
embolization device
12
embolization sites
12
peri-procedural occlusion
12
freedom device-related
12
first-in-human study
8
caterpillar™ arterial
8
arterial embolization
8
device-related sae
8
embolization
7

Similar Publications