Background: Transsulcal minimally invasive parafasicular (TsMIP) approaches to brain tumor resection use tubular retractors to minimize iatrogenic brain injury. Dynamic cortical and subcortical continuous neurophysiological mapping facilitates safer resection of motor-eloquent tumors.

Objective: To describe a new technique to address the challenge of combining TsMIP with tubular retractors and dynamic subcortical mapping using a single electrified stimulating microdebrider instrument.

Methods: We adapted the NICO Myriad microdebrider with continuous monopolar stimulation electrification using high-frequency stimulation with the train-of-5 technique. We performed continuous subcortical mapping using this device and compared it with standard dynamic monopolar subcortical mapping using a suction stimulation device. We found no significant difference in recorded stimulation response.

Results: Using a single operating instrument that provides synchronous tumor resection and monopolar subcortical mapping with the NICO Brainpath tubular retractor, we observed increased degrees of movement, faster surgical resection times with an enlarged working channel down the retractor, and improved safety because the stimulating probe sits 2 mm deep to the resection window.

Conclusion: We show that the adapted device is reliable and provides similar stimulation response as conventional subcortical mapping. We advocate the use of our adapted microdebrider in TsMIP tubular retractor approaches.

Download full-text PDF

Source
http://dx.doi.org/10.1227/ons.0000000000000281DOI Listing

Publication Analysis

Top Keywords

subcortical mapping
20
dynamic subcortical
8
tumor resection
8
tubular retractors
8
tsmip tubular
8
monopolar subcortical
8
tubular retractor
8
subcortical
7
mapping
7
tubular
5

Similar Publications

The lateral prefrontal cortex (LPFC) serves as a critical hub for higher-order cognitive and executive functions in the human brain, coordinating brain networks whose disruption has been implicated in many neurological and psychiatric disorders. While transcranial brain stimulation treatments often target the LPFC, our current understanding of connectivity profiles guiding these interventions based on electrophysiology remains limited. Here, we present a high-resolution probabilistic map of bidirectional effective connectivity between the LPFC and widespread cortical and subcortical regions.

View Article and Find Full Text PDF

Background: Frontotemporal dementia (FTD) encompasses diverse clinical phenotypes, primarily characterized by behavioral and/or language dysfunction. A newly characterized variant, semantic behavioral variant FTD (sbvFTD), exhibits predominant right temporal atrophy with features bridging behavioral variant FTD (bvFTD) and semantic variant primary progressive aphasia (svPPA). This study investigates the longitudinal structural MRI correlates of these FTD variants, focusing on cortical and subcortical structural damage to aid differential diagnosis and prognosis.

View Article and Find Full Text PDF

Iron-the most abundant magnetic brain substance-is essential for many biological processes, including dopamine and myelin synthesis. Quantitative susceptibility mapping (QSM) MRI has recently linked altered subcortical magnetic susceptibility (χ) to schizophrenia. Since χ is increased by iron and decreased by myelin, abnormal levels of either could underlie these QSM differences.

View Article and Find Full Text PDF

Decision making and learning processes together enable adaptive strategic behavior. Animal studies demonstrated the importance of subcortical regions in these cognitive processes, but the human subcortical contributions remain poorly characterized. Here, we study choice and learning processes in the human subcortex, using a tailored ultra-high field 7T functional MRI protocol combined with joint models.

View Article and Find Full Text PDF

Direct myelin imaging with inversion-recovery ultrashort-echo-time (IR-UTE) is highly motion-sensitive, yet extra hardware or longer scans are impractical. We evaluated whether a superior-inferior (SI) self-navigator with bit-reversed spoke-angles mitigates motion artifacts without extending acquisition. Dual-echo IR-UTE was implemented at 3T.

View Article and Find Full Text PDF