Article Abstract

Background: Selective internal radiotherapy based on transarterial radioembolization (TARE) with yttrium-90 ( Y) microspheres is an established treatment for primary or metastatic liver disease.

Purpose: The objective of this work is to optimize the dosimetry of patients treated with Y TARE, using positron emission tomography (PET) images.

Methods: The NEMA 2012 PET phantom was filled with nearly 3.9 GBq of Y activity and acquired at days 0, 3, 5, 7, and 9 on a classic time-of-flight PET/computed tomography (CT) scanner (Philips TF64) and on a silicon photomultiplier (SiPM)-based PET/CT scanner (Philips Vereos). Acquisitions were carried on following the guidelines proposed in a previously published multicentric trial and images were reconstructed by varying and combining the available parameters. Comparisons were performed to identify the best set(s) of parameters leading to the most accurate Y-PET image(s), in terms of activity distribution. Then, for both scanners, the best images were analyzed with Simplicit Y, a personalized dosimetry software using multicompartmental Medical Internal Radiation Dose model. The comparison between measured and true doses allowed to identify the image granting the most consistent dose estimations and, therefore, to designate the set of parameters to be applied on patients' data for the reconstruction of optimized clinical images. Posttreatment dosimetry of four patients was then realized with Simplicit Y using optimized imaging datasets.

Results: Based on activity distribution comparisons and dose estimations over phantom and patients data, the SiPM-based PET/CT system appeared more suitable than the photomultiplier tube-based TF64 for Y-PET imaging. With the SiPM-based PET/CT system, reconstructed images with a 2-mm voxel size combined with the application of the point spread function correction led to the most accurate results for quantitative Y measures.

Conclusions: For the SiPM-based PET/CT scanner, an optimized set of reconstruction parameters has been identified and applied on patients' data in order to generate the most accurate image to be used for an improved personalized Y-PET dosimetry, ensuring a reliable evaluation of the delivered doses.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.15880DOI Listing

Publication Analysis

Top Keywords

sipm-based pet/ct
16
dosimetry patients
8
scanner philips
8
pet/ct scanner
8
activity distribution
8
dose estimations
8
applied patients'
8
patients' data
8
pet/ct system
8
sipm-based
5

Similar Publications

Purpose: This study aimed to compare the image quality of the Siemens Biograph mCT40 (photomultiplier-based system - PMT) and the Siemens Vision600 (silicon photomultiplier-based system - SiPM) using a modified NEMA IEC Body phantom (Data Spectrum).

Methods: SiPM-based Vision600 has a smaller crystal size (3.2 × 3.

View Article and Find Full Text PDF

Purpose/aim: The increasing population age highlights the critical need for early brain disease diagnosis, especially in disorders such as dementia. Consequently, a notable focus has been on developing dedicated brain positron emission tomography (PET) scanners, which offer higher resolution and sensitivity than whole-body PET scanners. This study aims to design and performance evaluation of an LYSO-based dedicated brain PET scanner.

View Article and Find Full Text PDF

Objectives: We investigated image quality and standardized uptake values (SUVs) for different lesion sizes using clinical data generated by F-FDG-prone breast silicon photomultiplier (SiPM)-based positron emission tomography/computed tomography (PET/CT).

Methods: We evaluated the effect of point-spread function (PSF) modeling and Gaussian filtering (Gau) and determined the optimal reconstruction conditions. We compared the signal-to-noise ratio (SNR), contrast, %coefficient of variation (%CV), SUV, and Likert scale score between ordered-subset expectation maximization (OSEM) time-of-flight (TOF) and OSEM+TOF+PSF in phantom and clinical studies.

View Article and Find Full Text PDF

The Omni Legend 32 PET/CT system features silicon photomultiplier (SiPM)-based detectors with bismuth germanium oxide crystals and a 32-cm axial field of view (FOV). The present study aimed to determine the performance characteristics of the Omni Legend 32 PET/CT system according to National Electrical Manufacturers Association (NEMA) NU 2-2018 standards. The PET component of this system comprises 22 detector modules; each module contains 24 detector blocks with 72 bismuth germanium oxide crystals with a volume of 4.

View Article and Find Full Text PDF

Purpose: Sensitivity and count rate performance of the latest PET/CT scanners with a silicon photomultiplier (SiPM) have been substantially improved compared to scanners with a photomultiplier tube (PMT), thereby promising a low-dose whole-body PET scan with maintaining image quality. However, it is ethically difficult to verify the low-dose protocol in actual clinical settings. In this study, we investigated the effect of dose reduction on reconstructed images by using a low-dose simulation technique, i.

View Article and Find Full Text PDF