Article Abstract

Background: The recently emerged SARS-CoV-2 B.1.1.529 lineage and its sublineages (Omicron variant) pose a new challenge to healthcare systems worldwide due to its ability to efficiently spread in immunized populations and its resistance to currently available monoclonal antibody therapies. RT-PCR-based variant tests can be used to screen large sample-sets rapidly and accurately for relevant variants of concern (VOC). The aim of this study was to establish and validate a multiplex assay on the cobas 6800/8800 systems to allow discrimination between the two currently circulating VOCs, Omicron and Delta, in clinical samples.

Methods: Primers and probes were evaluated for multiplex compatibility. Analytic performance was assessed using cell culture supernatant of an Omicron variant isolate and a clinical Delta variant sample, normalized to WHO-Standard. Clinical performance of the multiplex assay was benchmarked against NGS results.

Results: In silico testing of all oligos showed no interactions with a high risk of primer-dimer formation or amplification of human DNA/RNA. Over 99.9% of all currently available Omicron variant sequences are a perfect match for at least one of the three Omicron targets included in the multiplex. Analytic sensitivity was determined as 19.0 IU/mL (CI95%: 12.9-132.2 IU/mL) for the A67V + del-HV69-70 target, 193.9 IU/mL (CI95%: 144.7-334.7 IU/mL) for the E484A target, 35.5 IU/mL (CI95%: 23.3-158.0 IU/mL) for the N679K + P681H target and 105.0 IU/mL (CI95%: 80.7-129.3 IU/mL) for the P681R target. All sequence variances were correctly detected in the clinical sample set (225/225 Targets).

Conclusion: RT-PCR-based variant screening compared to whole genome sequencing is both rapid and reliable in detecting relevant sequence variations in SARS-CoV-2 positive samples to exclude or verify relevant VOCs. This allows short-term decision-making, e.g., for patient treatment or public health measures.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8950896PMC
http://dx.doi.org/10.3390/v14030608DOI Listing

Publication Analysis

Top Keywords

iu/ml ci95%
16
omicron variant
12
sars-cov-2 b11529
8
rt-pcr-based variant
8
multiplex assay
8
iu/ml
8
omicron
6
variant
6
clinical
5
multiplex
5

Similar Publications

Background: The large dengue (DENV) and chikungunya (CHIKV) outbreaks observed during the last decade across the world, as well as local transmissions in non-endemic areas are a growing concern for blood safety. The aim of this study was to evaluate and compare the sensitivity of nucleic acid tests (NAT) detecting DENV and CHIKV RNA.

Materials And Methods: Using DENV 1 to 4 International Standards, the limits of detection (LODs) calculated by probit analysis of two NAT assays; the cobas CHIKV/DENV assay (Roche Diagnostics) and the Procleix Dengue Virus Assay (Grifols) were compared.

View Article and Find Full Text PDF

Serum Hepatitis B core-related antigen (HBcrAg) level moderately correlates with cccDNA. We examined whether HBcrAg can add value in monitoring the effect of peginterferon (PEG-IFN) therapy for HBeAg-negative chronic hepatitis B (CHB) infection. Thus, serum HBcrAg level was measured in 133 HBeAg-negative, mainly Caucasian CHB patients, treated with 48 weeks of PEG-IFN alfa-2a.

View Article and Find Full Text PDF

Background: Although certain HBV mutations are known to affect the expression of Hepatitis e antigen, their association with HBV viral level or clinical outcomes is less clear.

Objectives: We evaluated associations between different mutations in the Basal Core promoter (BCP) and Pre-core (PC) regions of HBV genome and subsequent changes in HBV viral DNA level over seven years in a population of untreated HBeAg negative chronic hepatitis B (CHB) participants in Northeast of Iran.

Materials And Methods: Participants in the current study were drawn from the Golestan Hepatitis B Cohort Study (GHBCS), a cohort of approximately 2590 HBsAg positive subjects (living in Gonbad city) embedded in the Golestan Cohort Study (GCS).

View Article and Find Full Text PDF

Background: The QuantiFERON®-TB Gold In-Tube test (QFT-GIT) detects Mycobacterium tuberculosis (Mtb) infection by measuring release of interferon gamma (IFN-γ) when T-cells (in heparinized whole blood) are stimulated with specific Mtb antigens. The amount of IFN-γ is determined by enzyme-linked immunosorbent assay (ELISA). Automation of the ELISA method may reduce variability.

View Article and Find Full Text PDF