A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Accurate Correction of the "Bulk Response" in Surface Plasmon Resonance Sensing Provides New Insights on Interactions Involving Lysozyme and Poly(ethylene glycol). | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Surface plasmon resonance is a very well-established surface sensitive technique for label-free analysis of biomolecular interactions, generating thousands of publications each year. An inconvenient effect that complicates interpretation of SPR results is the "bulk response" from molecules in solution, which generate signals without really binding to the surface. Here we present a physical model for determining the bulk response contribution and verify its accuracy. Our method does not require a reference channel or a separate surface region. We show that proper subtraction of the bulk response reveals an interaction between poly(ethylene glycol) brushes and the protein lysozyme at physiological conditions. Importantly, we also show that the bulk response correction method implemented in commercial instruments is not generally accurate. Using our method, the equilibrium affinity between polymer and protein is determined to be = 200 μM. One reason for the weak affinity is that the interaction is relatively short-lived (1/ < 30 s). Furthermore, we show that the bulk response correction also reveals the dynamics of self-interactions between lysozyme molecules on surfaces. Besides providing new insights on important biomolecular interactions, our method can be widely applied to improve the accuracy of SPR data generated by instruments worldwide.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9040059PMC
http://dx.doi.org/10.1021/acssensors.2c00273DOI Listing

Publication Analysis

Top Keywords

bulk response
16
"bulk response"
8
surface plasmon
8
plasmon resonance
8
polyethylene glycol
8
biomolecular interactions
8
response correction
8
surface
5
accurate correction
4
correction "bulk
4

Similar Publications