Article Abstract

Purpose: To evaluate whether the bony impingement lesion in elbow osteoarthritis can be removed accurately, as planned during arthroscopy, by using the computer-aided navigation system and performing mock surgery using 3-dimensional (3D)-printed bone models for clinical applications.

Methods: We performed mock surgery using 3D-printed plaster bone models of the humerus of 15 actual patients with elbow osteoarthritis. Two types of experiments were conducted to evaluate the surgical accuracy. Three surgeons performed the mock surgery, each with 15 bone models (total, 45 trials). Surgical accuracy was based on the mean of 45 trials. The differences in surgical accuracy among the 3 surgeons were also evaluated (mean 15 trials). The same surgeon performed 30 trials, and the difference in surgical accuracy between the first and the second halves was also evaluated (mean 15 trials).

Results: The spatial error in the entire elbow joint was 1.13 mm. In terms of resection volume, a mean of 8% more volume was resected than was planned, and 85% of the planned area was resected. In our experiments, the surgical accuracy was significantly lower in the anterior than in the posterior joint. Intrarater reliability was intraclass correlation (ICC) 0.81 and inter-rater reliability was ICC 0.87.

Conclusions: Surgery using computer-aided navigation systems for arthroscopic debridement of the elbow provided accuracy comparable to that in other joints.

Clinical Relevance: Arthroscopic debridement of elbow osteoarthritis requires advanced surgical skills because accurate identification of the bony impingement legion is difficult during surgery. Surgery using computer-aided navigation systems for arthroscopic debridement of the elbow will provide real-time tracking of both the surgical instruments and bony impingement lesions as well as solve the technical difficulties of arthroscopic surgery of the elbow joint.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8689228PMC
http://dx.doi.org/10.1016/j.asmr.2021.07.025DOI Listing

Publication Analysis

Top Keywords

surgical accuracy
20
arthroscopic debridement
16
debridement elbow
16
elbow osteoarthritis
16
computer-aided navigation
16
navigation systems
12
bony impingement
12
mock surgery
12
bone models
12
elbow
8

Similar Publications

Background: Breast-conserving surgery (BCS) is the primary surgical approach for patients with breast cancer. The accurate determination of surgical margins during BCS is critical for patient prognosis; however, time constraints and limitations in current pathological techniques often prevent pathologists from performing this assessment intraoperatively. The inability to reliably assess margins during surgery can lead to incomplete tumor removal and the need for additional surgeries.

View Article and Find Full Text PDF

Dural penetration patterns of the suboccipital vertebral arteries.

Ann Anat

September 2025

Department of Anatomy, School of Medicine, Faculty of Health Sciences, National and Kapodistrian University of Athens, Greece; "VARIANTIS" Research Laboratory, Department of Clinical Anatomy, Mazovian Academy in Plock, Poland.

Background: The vertebral artery (VA) undergoes a critical anatomical transition as it pierces the dura mater at the craniocervical junction. Precise knowledge of dural penetration patterns and angulation is essential for diagnostic imaging, neurosurgical planning, and minimizing iatrogenic risk in posterior fossa procedures.

Methods: This retrospective imaging study evaluated 100 adult patients who underwent 1.

View Article and Find Full Text PDF

Polyneuropathy in Kidney Transplant Recipients: Accuracy of a New Clinical Diagnostic Scoring System.

J Peripher Nerv Syst

September 2025

Department of Neurology, University Medical Center Groningen, University of Groningen, Groningen, the Netherlands.

Background And Aims: Polyneuropathy is highly prevalent among kidney transplant recipients (KTR), underscoring the need for an accurate yet easy-to-perform diagnostic method to improve understanding and enable early identification of treatable cases.

Methods: This study included KTR at least 12 months post-transplant at the University Medical Centre Groningen, the Netherlands. An expert panel assessed polyneuropathy through a structured neurological examination, quantitative sensory testing, and nerve conduction studies.

View Article and Find Full Text PDF

Multi-voxel pattern analysis of face and word encoding fMRI in people with temporal lobe epilepsy and healthy individuals.

Epilepsy Behav

September 2025

Department of Clinical and Experimental Epilepsy, University College London, London the United Kingdom of Great Britain and Northern Ireland; MRI Unit, Chalfont Centre for Epilepsy, Bucks, the United Kingdom of Great Britain and Northern Ireland. Electronic address:

Memory functional MRI (fMRI) has been used to explore cognitive processing in people with refractory temporal lobe epilepsy (TLE) to predict memory decline after anterior temporal lobe resection (ATLR). Traditional studies employed univariate analysis (UVA), focusing on isolated voxel activity in mesial temporal regions. By contrast, multivariate pattern analysis (MVPA), examines distributed activity patterns , offering deeper insight into neural networks supporting cognitive functions.

View Article and Find Full Text PDF

Background: The limited workspace and strong magnetic field inside MRI challenge the design of the prostate puncture robot. Simplifying the robot's structure is crucial.

Methods: This paper proposes a parallel cable-driven (PCD) prostate puncture robot, and conducts a preliminary material design.

View Article and Find Full Text PDF