Article Abstract

Purpose/objective: To evaluate intra-fraction target shift during automated mono-isocentric linac-based stereotactic radiosurgery with open-face mask system and optical real-time tracking.

Materials/methods: Ninety-five patients were treated using automated linac-based stereotactic radiosurgery in 1-5 fractions with single isocenter for a total of 195 fractions. During treatment, patient positioning was tracked real-time with optical surface guidance and immobilized with a rigid open-face mask. Patients were re-positioned if optical surface guidance error exceeded 1 mm magnitude or 1°. Translational and rotational intra-fractional changes were determined by post-treatment CBCT matched to the planning CT. Target specific error was calculated by translation and rotation matrices applied to isocenter and target spatial coordinates.

Results: For 132 fractions with isocenter within a single target, the median shift magnitude was 0.40 mm with a maximum shift of 1.17 mm. A total of 398 targets treated for plans having multiple or single targets that lied outside isocenter, resulted in a median shift magnitude of 0.46 mm, with median translational shifts of 0.20 mm and 0.20° rotational shifts. A 1 mm PTV margin was insufficient in 18% of targets at a distance greater than 6 cm away from isocenter, but sufficient for 96% of targets within 6 cm.

Conclusions: The findings of this study support 1 mm PTV expansion due to intra-fraction motion to ensure target coverage for plans with isocenter placement less than 6 cm away from the targets.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2021.11.012DOI Listing

Publication Analysis

Top Keywords

intra-fraction motion
8
automated linac-based
8
mask system
8
linac-based stereotactic
8
stereotactic radiosurgery
8
open-face mask
8
optical surface
8
surface guidance
8
median shift
8
shift magnitude
8

Similar Publications

Background: Intra-fraction motion management techniques, including beam gating and intra-fraction drift correction (IDC), have recently been introduced on the Unity MR-linac (Elekta AB, Stockholm, Sweden) to mitigate the dosimetric impact of motion during treatment. However, residual motion (e.g.

View Article and Find Full Text PDF

Background And Purpose: In pancreatic stereotactic body radiotherapy (SBRT), accurate motion management is crucial for the safe delivery of high doses per fraction. Intra-fraction tracking with magnetic resonance imaging-guidance for gated SBRT has shown potential for improved local control. Visualisation of pancreas (and surrounding organs) remains challenging in intra-fraction kilo-voltage (kV) imaging, requiring implanted fiducials.

View Article and Find Full Text PDF

A systematic review of the dosimetric consequences of the interplay effect.

Phys Med

June 2025

Radiotherapy Physics, Belfast City Hospital, Belfast Health & Social Care Trust, Belfast, UK; The Patrick G Johnston Centre for Cancer Research, Queen's University of Belfast, Belfast, UK.

Purpose: The presence of intrafraction motion introduces challenges to thoracic irradiation. The objective of this review is to evaluate the impact of intra-fraction motion on target dose in lung cancer (interplay effect).

Method: A comprehensive search was conducted in 2024 across three databases using predefined keywords related to the interplay effect in conjunction with lung cancer.

View Article and Find Full Text PDF

The objective of this work was to investigate the feasibility of using 0 mm PTV margin in online adaptive radiotherapy for the first fractions, in combination with treatment-specific local compensation of accumulated underdosage to the target in the last fraction.Intrafraction motion patterns and delineations of twelve patients with prostate cancer were selected to cover a range of observed systematic and random inter- and intrafraction motion patterns. Treatment plans with 0 and 3 mm margins were created and dose was accumulated rigidly using the observed motion patterns.

View Article and Find Full Text PDF

Purpose: MR-guided radiotherapy (MRgRT) on the MR-linac (MRL) with daily online plan adaptation enables better control of inter-fraction variability. Recently, Comprehensive Motion Management (CMM) was introduced for the 1.5T MRL.

View Article and Find Full Text PDF