Article Abstract

Purpose: The purpose of this study was to evaluate the accuracy of four-dimensional (4D) flow MRI for direct assessment of peak velocity, flow volume, and momentum of a mitral regurgitation (MR) flow jets using an in vitro pulsatile jet flow phantom. We systematically investigated the impact of spatial resolution and quantification location along the jet on flow quantities with Doppler ultrasound as a reference for peak velocity.

Methods: Four-dimensional flow MRI data of a pulsatile jet through a circular, elliptical, and 3D-printed patient-specific MR orifice model was acquired with varying spatial resolution (1.5-5 mm isotropic voxel). Flow rate and momentum of the jet were quantified at various axial distances (x = 0-50 mm) and integrated over time to calculate Vol and MTI . In vivo assessment of Vol and MTI was performed on 3 MR patients.

Results: Peak velocities were comparable to Doppler ultrasound (3% error, 1.5 mm voxel), but underestimated with decreasing spatial resolution (-40% error, 5 mm voxel). Vol was similar to regurgitant volume (RVol) within 5 mm, and then increased linearly with the axial distance (19%/cm) because of flow entrainment. MTI remained steady throughout the jet (2%/cm) as theoretically predicted. Four and 9 voxels across the jet were required to measure flow volume and momentum-time-integral within 10% error, respectively.

Conclusion: Four-dimensional flow MRI detected accurate peak velocity, flow rate, and momentum for in vitro MR-mimicking flow jets. Spatial resolution significantly impacted flow quantitation, which otherwise followed predictions of flow entrainment and momentum conservation. This study provides important preliminary information for accurate in vivo MR assessment using 4D flow MRI.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.29082DOI Listing

Publication Analysis

Top Keywords

flow mri
20
flow
17
spatial resolution
16
four-dimensional flow
12
regurgitation flow
8
mitral regurgitation
8
peak velocity
8
velocity flow
8
flow volume
8
flow jets
8

Similar Publications

Application of high-resolution cone-beam computed tomography for evaluation of endothelialization after flow diverter implantation for unruptured intracranial aneurysms.

AJNR Am J Neuroradiol

September 2025

From the Department of Interventional Radiology (Shuailong Shi, Shuhai Long, Ji Ma, Peijie Lu, Jie Yang, Ye Wang, Tengfei Li), The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China; The School of Life Science and Technology (Zhike Zhang), Xi'an Jiaotong University, Xi'an, China; an

Background And Purpose: Although high-resolution cone-beam computed tomography (HR-CBCT) is used for immediate evaluation of stent apposition, studies using this technique to evaluate flow diverter (FD) endothelialization during follow-up are limited. The study aims to investigate the potential of HR-CBCT in assessing FD endothelialization and identify factors influencing poor endothelialization.

Materials And Methods: The clinical and imaging data of patients with unruptured intracranial aneurysms (UIAs) treated by FDs from March 2019 to October 2023 were retrospectively analyzed.

View Article and Find Full Text PDF

Fetal 4D Flow CMR for Advanced Diagnostics of Congenital Heart Disease: A Prospective Cohort Study.

Eur Heart J Cardiovasc Imaging

September 2025

Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Venusberg-Campus 1, 53127, Bonn, Germany.

Aims: Fetal circulation undergoes complex changes in congenital heart disease (CHD) that are challenging to assess with fetal echocardiography. This study aimed to assess clinical feasibility and diagnostic value of 4D flow cardiac magnetic resonance (CMR) in fetal CHD.

Methods And Results: Pregnant women in advanced third trimester pregnancy with fetal CHD were prospectively recruited for fetal CMR between 08/2021 and 11/2024.

View Article and Find Full Text PDF

Full Free-Breathing Cardiac MRI: Enhancing Efficiency and Image Quality in Clinical Practice.

J Cardiovasc Magn Reson

September 2025

Department of Magnetic Resonance Imaging, Fuwai Hospital and National Center for Cardiovascular Diseases, Peking Union Medical College and Chinese Academy of Medical Sciences, Beijing 100037, China; Key Laboratory of Cardiovascular Imaging, Chinese Academy of Medical Sciences, Beijing 100730, China.

Background: Conventional cardiac magnetic resonance (CMR) examinations require patients to repeatedly hold their breath, which can reduce examination efficiency and pose challenges for patients unable to do so. This study aimed to demonstrate the feasibility and effectiveness of a full free-breathing CMR protocol in clinical practice.

Methods: Patients prospectively enrolled in this study underwent a full free-breathing CMR exam on a 3T scanner between June 1 and June 30, 2024.

View Article and Find Full Text PDF

Background: Azygos vein aneurysm (AVA) is a rare thoracic pathology that is frequently misdiagnosed. While contrast-enhanced chest computed tomography (CT) or magnetic resonance imaging (MRI) can delineate AVA location and size, these techniques lack the capability for dynamic real-time assessment of internal architecture.

Case Presentation: We present a highly unusual case of a 67-year-old woman who had an isolated azygos vein aneurysm presenting with dysphagia.

View Article and Find Full Text PDF

Glioblastoma is characterized by aggressive infiltration into surrounding brain tissue, hindering complete surgical resection and contributing to poor patient outcomes. Identifying tumor-specific invasion patterns is essential for advancing our understanding of glioblastoma progression and improving surgical and radiotherapeutic strategies. Here, we leverage in vivo dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) to noninvasively quantify interstitial fluid velocity, direction, and diffusion within and around glioblastomas.

View Article and Find Full Text PDF