Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Quantitative cardiovascular magnetic resonance T1-mapping is increasingly used for myocardial tissue characterization. However, the lack of standardization limits direct comparability between centers and wider roll-out for clinical use or trials.

Purpose: To develop a quality assurance (QA) program assuring standardized T1 measurements for clinical use.

Methods: MR phantoms manufactured in 2013 were distributed, including ShMOLLI T1-mapping and reference T1 and T2 protocols. We first studied the T1 and T2 dependency on temperature and phantom aging using phantom datasets from a single site over 4 years. Based on this, we developed a multiparametric QA model, which was then applied to 78 scans from 28 other multi-national sites.

Results: T1 temperature sensitivity followed a second-order polynomial to baseline T1 values (R > 0.996). Some phantoms showed aging effects, where T1 drifted up to 49% over 40 months. The correlation model based on reference T1 and T2, developed on 1004 dedicated phantom scans, predicted ShMOLLI-T1 with high consistency (coefficient of variation 1.54%), and was robust to temperature variations and phantom aging. Using the 95% confidence interval of the correlation model residuals as the tolerance range, we analyzed 390 ShMOLLI T1-maps and confirmed accurate sequence deployment in 90%(70/78) of QA scans across 28 multiple centers, and categorized the rest with specific remedial actions.

Conclusions: The proposed phantom QA for T1-mapping can assure correct method implementation and protocol adherence, and is robust to temperature variation and phantom aging. This QA program circumvents the need of frequent phantom replacements, and can be readily deployed in multicenter trials.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7994017PMC
http://dx.doi.org/10.1016/j.ijcard.2021.01.026DOI Listing

Publication Analysis

Top Keywords

phantom aging
12
quality assurance
8
phantom
8
correlation model
8
robust temperature
8
assurance quantitative
4
quantitative cardiac
4
t1-mapping
4
cardiac t1-mapping
4
t1-mapping multicenter
4

Similar Publications

Unlabelled: Accurate assessment of proximal humeral volumetric bone mineral density (vBMD) is essential for surgical planning in shoulder pathology. However, age-related changes in proximal humeral vBMD remain poorly characterized. This study developed a deep learning-based method to assess proximal humeral vBMD and identified sex-specific age-related changes.

View Article and Find Full Text PDF

Background: Scaphoid fractures predominantly occur in young males, while distal radius fractures occur in elderly females, despite similar injury mechanisms. We hypothesized that age-related bone mineral density (BMD) changes differ between these bones, explaining their distinct fracture patterns. This study compared age-related BMD changes between the scaphoid and distal radius using quantitative computed tomography (qCT).

View Article and Find Full Text PDF

Objective: This study evaluates cementation gap, cement volume, and porosity levels in three-dimensional crown restorations using four permanent resin materials and printers, and examines changes after thermomechanical ageing.

Materials And Methods: In the study, phantom models prepared on the maxillary first premolar were used, and the digital scanning, CAD design, 3D printing, cementation, and ageing processes were carried out according to standard protocols. The samples were divided into four groups based on the permanent resin material used: Group C (Methacrylate-based resin), Group S (UDMA-based composite), Group P (Bis-GMA/TEGDMA-based), and Group V (Bis-EMA/UDMA-based).

View Article and Find Full Text PDF

Background: The increasing use of technologies operating between 10 and 200 kHz, such as RFID, wireless power transfer systems, and induction cooktops, raises concerns about electromagnetic interference (EMI) with cardiac implantable electronic devices (CIEDs). The mechanisms of interaction within this frequency range have been only partially addressed by both the scientific and regulatory communities.

Methods: A physical twin of a pacemaker/implantable defibrillator (PM/ICD) was developed to experimentally assess voltages induced at the input stage by low-to-mid-frequency magnetic fields.

View Article and Find Full Text PDF

We present the case of a 74-year-old woman with behavioral variant frontotemporal dementia (bvFTD) linked to a pathogenic TANK-binding kinase 1 (TBK1) mutation (c.1349_1352del; p.Ile450Lysfs*15).

View Article and Find Full Text PDF