Complex confounder-corrected R2* mapping for liver iron quantification with MRI.

Eur Radiol

Department of Radiology, University of Wisconsin-Madison, Wisconsin Institutes for Medical Research, Rm 2474, 1111 Highland Ave, Madison, WI, 53705, USA.

Published: January 2021


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objectives: MRI-based R2* mapping may enable reliable and rapid quantification of liver iron concentration (LIC). However, the performance and reproducibility of R2* across acquisition protocols remain unknown. Therefore, the objective of this work was to evaluate the performance and reproducibility of complex confounder-corrected R2* across acquisition protocols, at both 1.5 T and 3.0 T.

Methods: In this prospective study, 40 patients with suspected iron overload and 10 healthy controls were recruited with IRB approval and informed written consent and imaged at both 1.5 T and 3.0 T. For each subject, acquisitions included four different R2* mapping protocols at each field strength, and an FDA-approved R2-based method performed at 1.5 T as a reference for LIC. R2* maps were reconstructed from the complex data acquisitions including correction for noise effects and fat signal. For each subject, field strength, and R2* acquisition, R2* measurements were performed in each of the nine liver Couinaud segments and the spleen. R2* measurements were compared across protocols and field strength (1.5 T and 3.0 T), and R2* was calibrated to LIC for each acquisition and field strength.

Results: R2* demonstrated high reproducibility across acquisition protocols (p > 0.05 for 96/108 pairwise comparisons across 2 field strengths and 9 liver segments, ICC > 0.91 for each field strength/segment combination) and high predictive ability (AUC > 0.95 for four clinically relevant LIC thresholds). Calibration of R2* to LIC was LIC = - 0.04 + 2.62 × 10 R2* at 1.5 T and LIC = 0.00 + 1.41 × 10 R2* at 3.0 T.

Conclusions: Complex confounder-corrected R2* mapping enables LIC quantification with high reproducibility across acquisition protocols, at both 1.5 T and 3.0 T.

Key Points: • Confounder-corrected R2* of the liver provides reproducible R2* across acquisition protocols, including different spatial resolutions, echo times, and slice orientations, at both 1.5 T and 3.0 T. • For all acquisition protocols, high correlation with R2-based liver iron concentration (LIC) quantification was observed. • The calibration between confounder-corrected R2* and LIC, at both 1.5 T and 3.0 T, is determined in this study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7755713PMC
http://dx.doi.org/10.1007/s00330-020-07123-xDOI Listing

Publication Analysis

Top Keywords

acquisition protocols
24
confounder-corrected r2*
20
r2*
17
r2* mapping
16
15 t 30 t
16
complex confounder-corrected
12
liver iron
12
r2* acquisition
12
field strength
12
iron concentration
8

Similar Publications

Introduction: Medical physicists play a critical role in ensuring image quality and patient safety, but their routine evaluations are limited in scope and frequency compared to the breadth of clinical imaging practices. An electronic radiologist feedback system can augment medical physics oversight for quality improvement. This work presents a novel quality feedback system integrated into the Epic electronic medical record (EMR) at a university hospital system, designed to facilitate feedback from radiologists to medical physicists and technologist leaders.

View Article and Find Full Text PDF

3D isotropic FastView MRI localizer allows reliable torsion measurements of the lower limb.

Eur Radiol Exp

September 2025

Department of Orthopaedics and Trauma Surgery, Orthopaedic Oncology, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Munich, Germany.

Computed tomography (CT) and magnetic resonance imaging (MRI) are commonly used to assess femoral and tibial torsion. While CT offers high spatial resolution, it involves ionizing radiation. MRI avoids radiation but requires multiple sequences and extended acquisition time.

View Article and Find Full Text PDF

Fetal 4D Flow CMR for Advanced Diagnostics of Congenital Heart Disease: A Prospective Cohort Study.

Eur Heart J Cardiovasc Imaging

September 2025

Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Venusberg-Campus 1, 53127, Bonn, Germany.

Aims: Fetal circulation undergoes complex changes in congenital heart disease (CHD) that are challenging to assess with fetal echocardiography. This study aimed to assess clinical feasibility and diagnostic value of 4D flow cardiac magnetic resonance (CMR) in fetal CHD.

Methods And Results: Pregnant women in advanced third trimester pregnancy with fetal CHD were prospectively recruited for fetal CMR between 08/2021 and 11/2024.

View Article and Find Full Text PDF

Tracking the translocation of fluorescent-based reporters at the single-cell level in living mouse embryos requires specialized expertise in mouse embryology and deep computational skills. Here, we detail an approach to quantify cyclin-dependent kinase (CDK) activity levels in single cells throughout different stages of the pre-implantation embryo. We discuss in vitro culture strategies that enable efficient live fluorescent confocal image acquisition and subsequent cell tracking.

View Article and Find Full Text PDF

Background: An increasing amount of data has been acquired on the close relationship between Vitamin K intake and the progression of periodontitis. The purpose of this study was to estimate whether Vitamin K intake is associated with periodontitis among older adults in America using information from the National Health and Nutrition Examination Survey (NHANES) database.

Methods: A cross-sectional analysis of data on periodontal indices and vitamin K intake among American adults aged 60 years and older in the 2009-2014 NHANES database was extracted.

View Article and Find Full Text PDF