A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Dynamic early identification of hip replacement implants with high revision rates. Study based on the NJR data from UK during 2004-2012. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Hip replacement and hip resurfacing are common surgical procedures with an estimated risk of revision of 4% over 10 year period. Approximately 58% of hip replacements will last 25 years. Some implants have higher revision rates and early identification of poorly performing hip replacement implant brands and cup/head brand combinations is vital.

Aims: Development of a dynamic monitoring method for the revision rates of hip implants.

Methods: Data on the outcomes following the hip replacement surgery between 2004 and 2012 was obtained from the National Joint Register (NJR) in the UK. A novel dynamic algorithm based on the CUmulative SUM (CUSUM) methodology with adjustment for casemix and random frailty for an operating unit was developed and implemented to monitor the revision rates over time. The Benjamini-Hochberg FDR method was used to adjust for multiple testing of numerous hip replacement implant brands and cup/ head combinations at each time point.

Results: Three poorly performing cup brands and two cup/ head brand combinations have been detected. Wright Medical UK Ltd Conserve Plus Resurfacing Cup (cup o), DePuy ASR Resurfacing Cup (cup e), and Endo Plus (UK) Limited EP-Fit Plus Polyethylene cup (cup g) showed stable multiple alarms over the period of a year or longer. An addition of a random frailty term did not change the list of underperforming components. The model with added random effect was more conservative, showing less and more delayed alarms.

Conclusions: Our new algorithm is an efficient method for early detection of poorly performing components in hip replacement surgery. It can also be used for similar tasks of dynamic quality monitoring in healthcare.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7402470PMC
http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0236701PLOS

Publication Analysis

Top Keywords

hip replacement
24
revision rates
16
cup cup
12
hip
9
early identification
8
replacement implant
8
implant brands
8
brand combinations
8
replacement surgery
8
random frailty
8

Similar Publications