A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Verification of dose distribution in high-dose-rate brachytherapy using a nanoclay-based radio-fluorogenic gel dosimeter. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Dose distributions have become more complex with the introduction of image-guided brachytherapy in high-dose-rate (HDR) brachytherapy treatments. Therefore, to correctly execute HDR, conducting a quality assurance programme for the remote after-loading system and verifying the dose distribution in the patient treatment plan are necessary. The characteristics of the dose distribution of HDR brachytherapy are that the dose is high near the source and rapidly drops when the distance from the source increases. Therefore, a measurement tool corresponding to the characteristic is required. In this study, using an Iridium-192 (Ir-192) source, we evaluated the basic characteristics of a nanoclay-based radio-fluorogenic gel (NC-RFG) dosimeter that is a fluorescent gel dosimeter using dihydrorhodamine 123 hydrochloride as a fluorescent probe. The two-dimensional dose distribution measurements were performed at multiple source positions to simulate a clinical plan. Fluorescence images of the irradiated NC-RFG were obtained at a high resolution (0.04 mm pixel) using a gel scanner with excitation at 465 nm. Good linearity was confirmed up to a dose range of 100 Gy without dose rate dependence. The dose distribution measurement at the five-point source position showed good agreement with the treatment planning system calculation. The pass ratio by gamma analysis was 92.1% with a 2%/1 mm criterion. The NC-RFG dosimeter demonstrates to have the potential of being a useful tool for quality assurance of the dose distribution delivered by HDR brachytherapy. Moreover, compared with conventional gel dosimeters such as polymer gel and Fricke gel dosimeters it solves the problems of diffusion, dose rate dependence and inhibition of oxygen-induced reactions. Furthermore, it facilitates dose data to be read in a short time after irradiation, which is useful for clinical use.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/ab98d2DOI Listing

Publication Analysis

Top Keywords

dose distribution
24
hdr brachytherapy
12
dose
11
nanoclay-based radio-fluorogenic
8
radio-fluorogenic gel
8
gel dosimeter
8
quality assurance
8
nc-rfg dosimeter
8
dose rate
8
rate dependence
8

Similar Publications