Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Given poor outcomes, strategies to improve ventricular assist devices (VADs) for single-ventricle patients with bidirectional Glenn (BDG) palliation are needed.

Methods: This retrospective review describes an institutional experience with VAD support for patients with BDG from April 2011 to January 2019. Surgical strategies, complications, and causes of death are described. Survival to heart transplantation for various strategies are compared.

Results: A total of 7 patients with BDG (weights, 5.6 to 28.8 kg; ages, 7 months to 11 years) underwent VAD implantation. Three patients underwent implantation of Berlin Heart EXCOR devices (Berlin Heart, Inc, Spring, TX), 2 had HeartWare HVADs (Medtronic, Minneapolis, MN) implanted, and 2 patients underwent implantation of paracorporeal continuous flow devices. Four patients underwent ventricular inflow cannulation, and 3 underwent atrial inflow cannulation. At the time of VAD implantation, the BDG was left intact in 3 patients, taken down in 3 patients, and created de novo in 1 patient. Over a total of 420 VAD support days, 2 patients survived to heart transplantation, 1 patient with HeartWare ventricular cannulation and intact BDG (after 174 days) and another with Berlin Heart atrial cannulation and BDG take-down (after 72 days). There were 3 deaths within 2 weeks of VAD implantation (2 from respiratory failure, 1 from infection) and 2 deaths after 30 days as a result of strokes.

Conclusions: The surgical strategy and postoperative management of VAD with BDG are still evolving. Successful support can be achieved with (1) both pulsatile and continuous flow pumps, (2) atrial or ventricular cannulation, and (3) with or without BDG take-down. Surgical strategy should be determined by individual patient anatomy, physiology, and condition.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.athoracsur.2019.12.088DOI Listing

Publication Analysis

Top Keywords

vad implantation
12
patients underwent
12
berlin heart
12
patients
9
ventricular assist
8
bidirectional glenn
8
bdg
8
vad support
8
patients bdg
8
heart transplantation
8

Similar Publications

Following ventricular assist device (VAD) placement, families and clinicians often have differing perspectives. When adverse events reduce patients' quality of life, families and clinicians question the desirability of continuing VAD support. Given the increasing use of VAD in pediatrics, pediatric-specific guidelines for the process of compassionate deactivation (CD) of VAD are needed, based in part on the perspectives of pediatric heart failure clinicians.

View Article and Find Full Text PDF

Right heart failure (RHF) is a major cause of mortality in patients using left ventricular assist devices (LVADs). Vericiguat, an oral soluble guanylate cyclase stimulator, has recently been used to treat chronic heart failure (HF) with a reduced ejection fraction (EF) and is expected to improve RHF owing to its pulmonary vasodilatory and inotropic properties in patients with an LVAD. A 62 year old woman with advanced HF due to cardiac sarcoidosis was transferred to our center with HF recurrence and left ventricular (LV) thrombosis.

View Article and Find Full Text PDF

Background: With advances in respiratory care allowing for improved survival, cardiomyopathy has emerged as the leading cause of death in patients with Duchenne Muscular Dystrophy (DMD). As end-stage heart failure emerges as the primary life-limiting complication in DMD patients, their consideration for advanced heart failure therapies, including ventricular assist devices (VADs) and/or heart transplantation (HT), is increasing. To date, however, there are few published reports of HT in DMD patients.

View Article and Find Full Text PDF

Background: Living myocardial slices (LMS) are an emerging translational model for studying myocardial function, disease mechanisms, and therapeutics. However, the extent to which findings correlate with clinical characteristics is unknown. This study aimed to evaluate whether LMS retain patient-specific functional and pathological characteristics, reflecting diverse etiologies, pharmacological regimens, and clinical interventions.

View Article and Find Full Text PDF

The transcatheter EVOQUE tricuspid valve replacement system is an approved device for the treatment of patients with symptomatic severe tricuspid regurgitation despite medical therapy both in the United States and in Europe. Specific imaging requirements are needed in the screening and in the intraprocedural phase as well as to assess this valve after implantation, in order to assess its positioning and performance. In this review, focus will be on the postoperative echocardiographic assessment of device positioning and stability; evaluation of residual central tricuspid regurgitation or paravalvular leak; valve thrombosis/degeneration; interaction of the device frame with the right ventricle and adjacent structures, including the aortic valve and interventricular septum; and placement of pacemaker leads.

View Article and Find Full Text PDF