Purpose: To assess surface characteristics and implant stability after implantoplasty performed by two different instrument sequences regarding material loss, surface roughness, and fracture load resistance. Additionally, operators' subjective experience during instrumentation and the damage to neighboring teeth were evaluated.

Materials And Methods: Titanium implants were placed in the position of both first maxillary molars in models exposing 6 mm of their surface. Implantoplasty was performed in phantom heads: Exposed surfaces were instrumented with diamonds and Arkansas stones or abrasive stones and silicone polishers. Operators reported on abrasion, gloss, effectiveness, and tactility using a visual analog scale (VAS). Residual wall thickness of implants was measured on radiographs, material abrasion using three-dimensional (3D) scans, and surface roughness by contact profilometry. Maximum bending moments were measured.

Results: Residual thickness and weight loss were comparable after both treatments (0.3 ± 0.1 and 0.25 ± 0.07 mm and 0.22 ± 0.01 g, and 0.03 ± 0.01 mm and 0.02 ± 0.01 g, respectively, P > .05). Mean surface roughness was lower (P = .0001) for the group with the silicone polishers (0.4 ± 0.2 μm) compared with the group employing diamonds (0.8 ± 0.1 μm). Maximum bending moments showed neither intergroup differences nor stability loss compared with untreated implants. The stone-and-silicone polisher group showed less abrasion (4.6 ± 2.2) and higher gloss values (8.1 ± 1.4) than the diamond-and-Arkansas group (3.1 ± 1.3 and 4.1 ± 2.1, respectively). Superficial tooth injuries at proximal neighbor teeth were common (73% and 80%).

Conclusion: Implantoplasty did not weaken implant stability. The use of silicone polishers revealed lower surface roughness. Regarding surface smoothness, the instrumentation sequence employing silicon carbide and Arkansas stones followed by silicone polishers seems to be superior to the combination of diamond and Arkansas stones.

Download full-text PDF

Source
http://dx.doi.org/10.11607/jomi.7410DOI Listing

Publication Analysis

Top Keywords

surface roughness
16
silicone polishers
16
arkansas stones
12
implant stability
8
implantoplasty performed
8
stones silicone
8
maximum bending
8
bending moments
8
surface
7
titanium implant
4

Similar Publications

Cutibacterium acnes (C. acnes, formerly classified as Propionibacterium acnes) is a Gram-positive bacterium that contributes to the development of acne vulgaris, resulting in inflammation and pustule formation on the skin. In this study, we developed and synthesized a series of antimicrobial peptides (AMPs) that are derived from the skin secretion of Rana chensinensis.

View Article and Find Full Text PDF

Application of lignin extracted from fibers and aminated lignin in anionic dyes contaminated water remediation.

Int J Phytoremediation

September 2025

Department of Fashion and Textile Design, College of Arts and Design, Princess Nourah bint Abdulrahman University, Riyadh, Saudi Arabia.

In this paper, lignin was chemically extracted from fibers and modified with branched polyethyleneimine (BPEI) and the resulting samples were applied for the adsorption of two anionic dyes; Acid red 183 (AR183) and Acid blue 25 (AB25) from aqueous suspension. Analytical characterization methods including SEM, FT-IR, TGA/DTG, and XRD were used to analyze the studied samples. The images of the extracted lignin displayed a rough feature.

View Article and Find Full Text PDF

Purpose: To investigate the short-term impact of exposure to smoke from vegetation burns on ocular surface symptoms and signs.

Methods: Woody bushfuels were burnt in an enclosed room (Flammability Laboratory, University of Tasmania, Australia) to generate particulate matter and monitored in real time (Dust Trak II). Eighteen participants (aged 20-63 years, 8 males and 10 females) fitted with respirators were seated 1.

View Article and Find Full Text PDF

[Experimental optimization of paraffin sectioning techniques for the eyeball].

Zhonghua Yan Ke Za Zhi

September 2025

Ophthalmology Medical Center, The First Affiliated Hospital of Chongqing Medical University, Chongqing Key Laboratory for the Prevention and Treatment of Major Blinding Eye Diseases, Chongqing Branch (Municipality Division) of National Clinical Research Centre for Ocular Diseases, Chongqing 400016,

To explore optimized protocols for paraffin section preparation of the eyeball to enhance the histological visualization of key ocular structures. It was an experimental research, conducted from September 2022 to September 2024. The first experiment involved 18 porcine eyeballs, which were divided into five groups (six subgroups) by the random number table method.

View Article and Find Full Text PDF

The study explored HSPiP and QbD-(quality by design) enabled optimized cubosomes for sustained drug release, improved permeation, and enhanced oral bioavailability. OCUB1 (the optimized product) was characterized for size, zeta potential (ZP), thermal analysis, and surface roughness. drug release and hemolysis studies were carried out using a dialysis membrane and rat erythrocytes (4 % suspension), respectively.

View Article and Find Full Text PDF