Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
The BioFire FilmArray Respiratory Panel (FA RP) is a rapid multiplexed molecular assay approved for detection of viral and atypical bacterial pathogens in nasopharyngeal specimens. This study aimed to evaluate the performance of the BioFire FilmArray Respiratory Panel v1.7 on bronchoscopy specimens. We tested 133 bronchial specimens (87 archived and 46 prospectively collected) with the FA RP and compared the results to the Luminex NxTAG Respiratory Pathogen Panel (NxTAG RPP). After discordant analysis, 123 specimens gave concordant results using the FA RP and the NxTAG RPP for an overall agreement of 93.9% (kappa = 0.88 [95% CI 0.80-0.96]), a positive percent agreement of 93.7% (95% CI 83.7-97.7) and a negative percent agreement of 94.1% (95% CI 84.9-98.1). In conclusion, the BioFire FilmArray RP performed reliably to detect a broad range of respiratory pathogens in bronchoscopy specimens.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7132745 | PMC |
http://dx.doi.org/10.1016/j.diagmicrobio.2019.114880 | DOI Listing |