Comparison of vibrotactile and joint-torque feedback in a myoelectric upper-limb prosthesis.

J Neuroeng Rehabil

Department of Mechanical Engineering, Johns Hopkins University, 3101 Wyman Park Dr., Baltimore, 21218, USA.

Published: June 2019


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Despite the technological advancements in myoelectric prostheses, body-powered prostheses remain a popular choice for amputees, in part due to the natural sensory advantage they provide. Research on haptic feedback in myoelectric prostheses has delivered mixed results. Furthermore, there is limited research comparing various haptic feedback modalities in myoelectric prostheses. In this paper, we present a comparison of the feedback intrinsically present in body-powered prostheses (joint-torque feedback) to a commonly proposed feedback modality for myoelectric prostheses (vibrotactile feedback). In so doing, we seek to understand whether the advantages of kinesthetic feedback present in body-powered prostheses translate to myoelectric prostheses, and whether there are differences between kinesthetic and cutaneous feedback in prosthetic applications.

Methods: We developed an experimental testbed that features a cable-driven, voluntary-closing 1-DoF prosthesis, a capstan-driven elbow exoskeleton, and a vibrotactile actuation unit. The system can present grip force to users as either a flexion moment about the elbow or vibration on the wrist. To provide an equal comparison of joint-torque and vibrotactile feedback, a stimulus intensity matching scheme was utilized. Non-amputee participants (n=12) were asked to discriminate objects of varying stiffness with the prosthesis in three conditions: no haptic feedback, vibrotactile feedback, and joint-torque feedback.

Results: Results indicate that haptic feedback increased discrimination accuracy over no haptic feedback, but the difference between joint-torque feedback and vibrotactile feedback was not significant. In addition, our results highlight nuanced differences in performance depending on the objects' stiffness, and suggest that participants likely pay less attention to incidental cues with the addition of haptic feedback.

Conclusion: Even when haptic feedback is not modality matched to the task, such as in the case of vibrotactile feedback, performance with a myoelectric prosthesis can improve significantly. This implies it is possible to achieve the same benefits with vibrotactile feedback, which is cheaper and easier to implement than other forms of feedback.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6558922PMC
http://dx.doi.org/10.1186/s12984-019-0545-5DOI Listing

Publication Analysis

Top Keywords

haptic feedback
24
vibrotactile feedback
24
feedback
20
myoelectric prostheses
20
joint-torque feedback
12
body-powered prostheses
12
feedback myoelectric
8
prostheses
8
feedback modality
8
feedback vibrotactile
8

Similar Publications

Telerobotic surgery: a comprehensive two-decade evolution and the integration of emerging technologies.

Int J Surg

September 2025

Department of Gastrointestinal Surgery, Xiang'an Hospital of Xiamen University, School of Medicine, Xiamen University, Xiamen, People's Republic of China.

Telerobotic surgery has undergone remarkable advances over the past two decades, driven by the integration of sophisticated robotic platforms and modern communication technologies, thereby alleviating many constraints of conventional surgical procedures. Building upon previous studies that focused on individual specialties or specific innovations, this review provides a comprehensive and integrated perspective by tracing the evolution of the field and highlighting applications in gastroenterology, urology, neurology, and cardiology. Landmark achievements include the 2001 transatlantic remote cholecystectomy and the introduction of next-generation platforms such as the Hinotori surgical robot, enabling low-latency remote interventions.

View Article and Find Full Text PDF

Objectives: Teaching dental anesthesia techniques poses a considerable challenge, primarily due to the limited availability of tools that effectively replicate clinical procedures in preclinical settings. Over the past decade, haptic dental simulators have emerged as promising training aids for various dental procedures, including local anesthesia. This study aimed to evaluate the educational value of a haptic dental simulator in teaching the inferior alveolar nerve block (IANB) technique by assessing the experiences and perceptions of dental students with varying levels of clinical exposure.

View Article and Find Full Text PDF

Introduction/background: Virtual dental simulators with a haptics component have been used with great success in dental education for over a decade and is becoming an integral part of dental curricula. A large number of studies have been published about the users' perception, acceptance and attitude towards virtual dental simulators. However, no longitudinal or long-term studies to our knowledge have been conducted to evaluate the users' acceptance over time.

View Article and Find Full Text PDF

As the metaverse blurs the lines between physical and virtual realms, enhanced by haptic devices providing sensory feedback, it is poised to become integral to daily life. However, this new digital frontier is also a site for sexual violence. Women users increasingly report non-consensual touching, image-based sexual abuses and novel forms of gendered harm, often trivialised and inadequately addressed by current laws.

View Article and Find Full Text PDF

Both intracortical microstimulation (ICMS) and peripheral nerve stimulation (PNS) can restore tactile sensation to people living with physical disabilities, such as spinal cord injury (SCI) or amputation. While both techniques have demonstrated success in evoking meaningful sensations in the upper limb, they have only been investigated in separate studies with different patient populations, and thus their perceptual characteristics have never been systematically compared to determine the relative advantages and limitations of each approach. In this study, we directly compared the perceived sensations evoked by ICMS and PNS to those evoked by mechanical touch in a participant with sensory incomplete spinal cord injury.

View Article and Find Full Text PDF