A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1075
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3195
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Characteristics of hardware failure in patients undergoing surgical stabilization of rib fractures: A Chest Wall Injury Society multicenter study. | LitMetric

Characteristics of hardware failure in patients undergoing surgical stabilization of rib fractures: A Chest Wall Injury Society multicenter study.

J Trauma Acute Care Surg

From the Center for Trauma and Critical Care, Department of Surgery (B.S., R.A., P.G.), George Washington University, Washington, District of Columbia; Department of Surgery (F.M.P.), Denver General Hospital; Department of Surgery (A.R.D.), Baystate Health, Springfield, Massachusetts; Department of

Published: December 2019


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: Surgical stabilization of rib fractures (SSRF) is increasingly used for severe rib fractures/flail chest. There are no reports discussing mechanisms of failure of implanted hardware, its clinical presentation, or consequences. The purpose of this study was to evaluate the incidence, presenting signs, and clinical sequela of hardware failure after SSRF.

Methods: A multicenter, retrospective study was performed by a group of surgeons with a large SSRF case volume. All cases with known hardware failure from January 1, 2010, to December 31, 2017, were included. The surgeon's experience at the time of hardware implantation, specific implant used, number of failures the surgeon had experienced with the same system, and time from implantation to hardware failure were recorded. Additionally, patient demographics, including age, comorbid conditions, and number and location of rib fractures were recorded. Symptomatology associated with hardware failure and need for explant and/or reimplantation of hardware was also recorded. Nonparametric statistical tests were used to compare cohorts.

Results: Of 1,224 patients who underwent SSRF, 38 patients with 233 rib fractures and 279 fracture segments experienced hardware failure and were enrolled in the study. Twelve patients presented more than 3 months following injury. Median age was 54 years old and 34% were active smokers. One hundred forty-four plates were implanted with a median of four plates per patient. Median number of SSRF cases by each surgeon was 100 (range, 1-280). Fractures and hardware failure were most frequent in the anterolateral/lateral region. Hardware failure was mostly due to screw migration and plate fracture. Hardware failure was asymptomatic in 40% and presented as pain in 42% of cases. Fifty-five percent of the cases required explantation of hardware, and only 10% required SSRF again. There was no difference between the acute and chronic fracture cohorts.

Conclusion: Hardware failure after SSRF is rare and often asymptomatic. When present, it rarely requires redo SSRF.

Level Of Evidence: Therapeutic, level V.

Download full-text PDF

Source
http://dx.doi.org/10.1097/TA.0000000000002373DOI Listing

Publication Analysis

Top Keywords

hardware failure
40
rib fractures
16
hardware
13
failure
11
surgical stabilization
8
stabilization rib
8
ssrf
6
rib
5
fractures
5
characteristics hardware
4