Biosample Concentration Using Microscale Forward Osmosis with Electrochemical Monitoring.

Anal Chem

Department of Chemical Engineering , Northeastern University, Boston , Massachusetts 02115 , United States.

Published: June 2019


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

We report the design and operation of an integrated microfluidics system that uses cellulose ester dialysis membranes coupled with disposable carbon and copper electrodes for monitoring and concentration of microliter scale biofluid samples. Dialysis membranes are typically used for buffer exchange, but in this work, membranes with 100-500 Da MWCO were evaluated for feasibility in concentrating small volume samples. This is an alternative to the use of centrifugation, ultrafiltration, and evaporative methods, where quantitative inline monitoring of sample concentration is challenging. The impact of draw solution used, osmotic concentration gradient, pH, and temperature were studied for the optimized concentration of bodily fluids. A system using sucrose in the draw solution generated the best results, with water removal rates of 0.023 mL min. PBS, urine, and saliva samples were concentrated up to 20-fold (PBS), 15-fold (urine), and 5-fold (saliva) in less than 3 h. The osmotic system further showed a 5-fold increase in the electrochemical signal for detecting pyocyanin, a biomarker for early diagnostics of the Pseudomonas aeruginosa pathogen in urine and saliva samples. Overall, the osmotic system can be easily integrated with point of care diagnostic systems for low cost improvement in signal amplification and limit of detection.

Download full-text PDF

Source
http://dx.doi.org/10.1021/acs.analchem.9b02163DOI Listing

Publication Analysis

Top Keywords

dialysis membranes
8
draw solution
8
urine saliva
8
saliva samples
8
osmotic system
8
biosample concentration
4
concentration microscale
4
microscale forward
4
forward osmosis
4
osmosis electrochemical
4

Similar Publications

Long-distance aeromedical transport of critically ill patients is an increasingly important component of modern intensive care. However, the combination of veno-arterial extracorporeal membrane oxygenation (VA ECMO) and renal replacement therapy (RRT) during an intercontinental flight had never been previously documented. This case report describes the first known case of a 27 year old patient with fulminant viral myocarditis and multi-organ failure who was successfully repatriated from Bangkok (Thailand) to Paris (France) while receiving both VA ECMO and 6 hours of in-flight sustained low-efficiency dialysis (SLED).

View Article and Find Full Text PDF

Objectives: Severe obesity is an established risk factor for adverse cardiovascular events and heart transplantation (HT) outcomes in adults. However, the effect of severe obesity on children after HT is not well studied. We aimed to examine the prevalence and effect of severe obesity on pediatric HT.

View Article and Find Full Text PDF

Background: Delayed graft function is a common situation that leads to increased long-term rates of graft rejection and loss. It is seen increasingly more often, as the use of kidneys from donors after controlled cardiac death has become more widespread. This study aimed to identify factors contributing to its onset and determine how these factors may influence graft survival.

View Article and Find Full Text PDF

Rationale: Ureteral stricture is a complex urological condition often requiring surgical intervention. Autologous tissue grafts, such as lingual mucosa, have emerged as a promising option for reconstruction due to their favorable biocompatibility and vascularity. However, reports on complications associated with these techniques remain limited.

View Article and Find Full Text PDF

The study explored HSPiP and QbD-(quality by design) enabled optimized cubosomes for sustained drug release, improved permeation, and enhanced oral bioavailability. OCUB1 (the optimized product) was characterized for size, zeta potential (ZP), thermal analysis, and surface roughness. drug release and hemolysis studies were carried out using a dialysis membrane and rat erythrocytes (4 % suspension), respectively.

View Article and Find Full Text PDF