Purpose: A 16-channel receive (16Rx) radiofrequency (RF) array for 7T ultra-high field body MR imaging is presented. The coil is evaluated in conjunction with a 16-channel transmit/receive (16TxRx) coil and additionally with a 32-channel transmit/receive (32TxRx) remote body coil for RF transmit and serving as receive references.
Methods: The 16Rx array consists of 16 octagonal overlapping loops connected to custom-built detuning boards with preamplifiers. Performance metrics like noise correlation, g-factors, and signal-to-noise ratio gain were compared between 4 different RF coil configurations. In vivo body imaging was performed in volunteers using radiofrequency shimming, time interleaved acquisition of modes (TIAMO), and 2D spatially selective excitation using parallel transmit (pTx) in the spine.
Results: Lower g-factors were obtained when using the 16Rx coil in addition to the 16TxRx array coil configuration versus the 16TxRx array alone. Distinct signal-to-noise ratio gain using the 16Rx coil could be demonstrated in the spine region both for a comparison with the 16TxRx coil (>50% gain) in vivo and the 32TxRx coil (>240% gain) in a phantom. The 16Rx coil was successfully applied to improve anatomical imaging in the abdomen and 2D spatially selective excitation in the spine of volunteers.
Conclusion: The novel 16-channel Rx-array as an add-on to multichannel TxRx RF coil configurations provides increased signal-to-noise ratio, lower g-factors, and thus improves 7T ultra-high field body MR imaging.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mrm.27731 | DOI Listing |
MAGMA
September 2025
Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Introduction: This study explores high-impedance surface (HIS) metamaterial shields for enhancing the transmit field in whole-body MRI at 7 T. We studied the possibility of placing a metamaterial layer between the gradient coil and bore liner using electromagnetic simulations to evaluate B and SAR efficiency across different impedances.
Materials And Methods: Simulations were performed in three stages, first metamaterial design and characterization, then single-element dipole simulations with a homogenous phantom, and finally, simulations including a four-element arrays with a virtual body model, including the whole scanner geometry.
MAGMA
September 2025
Department of Medical Imaging, (766), Radboud University Medical Center, Geert Grooteplein 10Radboudumc, P.O. Box 9101, 6500 HB, Nijmegen, the Netherlands.
Objective: To improve B field homogeneity in prostate MR imaging and spectroscopy using a custom-designed 16-channel external local shim coil array.
Methods: In vivo prostate imaging was performed in seven healthy volunteers (mean age: 40.7 years) without bowel preparation.
Phys Med Biol
September 2025
Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen University Medical Faculty, Pauwelsstraße 20, Aachen, 52074, GERMANY.
Objective: Magnetic particle imaging (MPI) opens huge possibilities in image-guided therapy. Its effectiveness is strongly influenced by the quality of the magnetic nanoparticles (MNP) used as tracers. Besides MNP optimization following different synthesis routes, MNP assembly into linear structures can significantly enhance their performance in MPI.
View Article and Find Full Text PDFACS Appl Mater Interfaces
September 2025
School of Integrated Circuits, Shandong University, Jinan 250100, P. R. China.
Transient electronics that can degrade after fulfilling their designed functionalities offer transformative potentials in biomedical implants (eliminating secondary surgeries), ecofriendly consumer electronics (reducing e-waste), and secure systems. However, the development of reliable transient energy supplies remains a critical challenge, thus limiting their widespread implementation. Among various solutions, wireless power supplies via near-field inductive coupling stand out as particularly promising candidates.
View Article and Find Full Text PDFAbdom Radiol (NY)
September 2025
Peking University First Hospital, Beijing, China.
Purpose: To evaluate the feasibility and clinical utility of non-contrast magnetic resonance angiography (NC-MRA) using a modified balanced steady-state free precession (b-SSFP) technique combined with arterial spin labeling (ASL) for post-embolotherapy follow-up of renal artery aneurysms (RAAs), with digital subtraction angiography (DSA) as the reference standard.
Method: A total of 57 patients with RAAs underwent embolotherapy between June 2013 and July 2024. Among them, 15 RAAs from 14 patients underwent post-embolization surveillance with the NC-MRA and DSA.