Performance and Usability of Three Systems for Continuous Glucose Monitoring in Direct Comparison.

J Diabetes Sci Technol

2 Abteilung für Innere Medizin / Kompetenzzentrum Diabetes, Privatklinik Wehrle-Diakonissen, Salzburg, Austria.

Published: September 2019


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: To be able to compare continuous glucose monitoring (CGM) systems, they have to be worn in parallel by the same subjects. This study evaluated the performance and usability of three different CGM systems in direct comparison.

Method: In this open, prospective study at two sites, 54 patients with diabetes wore three CGM systems each (Dexcom G5™ Mobile CGM system [DG5], Guardian™ Connect system [GC], and a Roche CGM system [RCGM]) in parallel for 6 or 7 days in a mixed inpatient and outpatient setting. Capillary comparison measurements were performed using a self-monitoring of blood glucose (SMBG) system. During study site visits, glucose excursions were induced. Performance of the systems was evaluated by calculating mean absolute relative differences (MARD, calculated as absolute differences for glucose concentrations <100 mg/dL and as relative differences for glucose concentrations ≥100 mg/dL), and mean relative differences (MRD, bias) between CGM and SMBG results. In addition, usability of the systems was assessed.

Results: Overall MARD was 10.1 ± 2.1 for DG5, 11.5 ± 4.2 for GC, and 11.9 ± 5.6 for RCGM. Performance improved in all systems after the first day of use. All systems showed >99% of values within zones A and B of the consensus error grid. Overall, all CGM systems showed a small negative bias compared to SMBG. Usability of the systems differed regarding patch adhesion rate, failure rate, and patient rating. Most patients preferred GC, but in general all systems were rated positively.

Conclusion: All three CGM systems showed similar overall accuracy in this direct comparison, but small differences were observed with regard to specific glucose ranges and usability aspects.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6955463PMC
http://dx.doi.org/10.1177/1932296819826965DOI Listing

Publication Analysis

Top Keywords

cgm systems
20
three cgm
12
systems
9
performance usability
8
usability three
8
continuous glucose
8
glucose monitoring
8
direct comparison
8
cgm system
8
cgm
7

Similar Publications

Aims: To assess the frequency and management of hypoglycaemia during unstructured physical activity (PA) in adults with type 1 diabetes (T1D) using automated insulin delivery (AID) systems in real-life settings.

Materials And Methods: RAPPID is a prospective, multicenter, observational study conducted over 1 month in four French tertiary care centres. Adults with T1D using one of three AID systems (MiniMed 780G, Tandem t:slim X2 with Control-IQ, or Ypsopump with CamAPS FX) and performing ≥2 unstructured PA sessions per week were included.

View Article and Find Full Text PDF

Aims: This study aimed to assess the impact of the Omnipod 5 automated insulin delivery (AID) system on continuous glucose monitoring (CGM) metrics, HbA1c, and weight in a real-world setting. Additionally, independent predictors of glycaemic response were assessed.

Methods: Observational analysis of adults with type 1 diabetes using Omnipod 5 (n = 353).

View Article and Find Full Text PDF

Management of diabetes mellitus in hemodialysis is highly complex due to increased glycemic variability and hypoglycemic risk. The use of technologies applied to diabetes has been shown to improve glycemic control, however data in dialysis patients are limited. To describe the efficacy and safety of the minimed 780G AHCL system in a stable hemodialysis patient and during hospitalization in the Intensive Care Unit (ICU).

View Article and Find Full Text PDF

Implantable biosensors: Advancements and applications.

Prog Mol Biol Transl Sci

September 2025

Faculty of Dental Medicine and Oral Health Sciences, McGill University, Montreal, QC, Canada. Electronic address:

This chapter, "Implantable Biosensors: Advancements and Applications," provides a succinct overview of the state-of-the-art in implantable biosensor technology, highlighting both established clinical uses and promising areas of ongoing research. It begins by outlining the fundamental principles and advantages of these sensors, such as their precision in physiological monitoring and capability for real-time therapeutic interventions. A variety of implantable sensors are categorized, including biophysical and biochemical types, each designed for specific medical applications.

View Article and Find Full Text PDF

Background: Despite advances, glycemic control in people with type 2 diabetes (PwT2D) treated with oral antidiabetic medications (ADMs) often remains suboptimal. Continuous glucose monitoring (CGM) has shown promise in diabetes management, offering real-time insights into glucose trends. This study evaluates the impact of transitioning from conventional self-monitoring of blood glucose (SMBG) to CGM on glycemic outcomes and self-management in PwT2D receiving oral ADMs.

View Article and Find Full Text PDF