Background: To be able to compare continuous glucose monitoring (CGM) systems, they have to be worn in parallel by the same subjects. This study evaluated the performance and usability of three different CGM systems in direct comparison.
Method: In this open, prospective study at two sites, 54 patients with diabetes wore three CGM systems each (Dexcom G5™ Mobile CGM system [DG5], Guardian™ Connect system [GC], and a Roche CGM system [RCGM]) in parallel for 6 or 7 days in a mixed inpatient and outpatient setting. Capillary comparison measurements were performed using a self-monitoring of blood glucose (SMBG) system. During study site visits, glucose excursions were induced. Performance of the systems was evaluated by calculating mean absolute relative differences (MARD, calculated as absolute differences for glucose concentrations <100 mg/dL and as relative differences for glucose concentrations ≥100 mg/dL), and mean relative differences (MRD, bias) between CGM and SMBG results. In addition, usability of the systems was assessed.
Results: Overall MARD was 10.1 ± 2.1 for DG5, 11.5 ± 4.2 for GC, and 11.9 ± 5.6 for RCGM. Performance improved in all systems after the first day of use. All systems showed >99% of values within zones A and B of the consensus error grid. Overall, all CGM systems showed a small negative bias compared to SMBG. Usability of the systems differed regarding patch adhesion rate, failure rate, and patient rating. Most patients preferred GC, but in general all systems were rated positively.
Conclusion: All three CGM systems showed similar overall accuracy in this direct comparison, but small differences were observed with regard to specific glucose ranges and usability aspects.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6955463 | PMC |
http://dx.doi.org/10.1177/1932296819826965 | DOI Listing |
Diabetes Obes Metab
September 2025
Diabetes Care Unit, Nancy University Hospital, Nancy, France.
Aims: To assess the frequency and management of hypoglycaemia during unstructured physical activity (PA) in adults with type 1 diabetes (T1D) using automated insulin delivery (AID) systems in real-life settings.
Materials And Methods: RAPPID is a prospective, multicenter, observational study conducted over 1 month in four French tertiary care centres. Adults with T1D using one of three AID systems (MiniMed 780G, Tandem t:slim X2 with Control-IQ, or Ypsopump with CamAPS FX) and performing ≥2 unstructured PA sessions per week were included.
Diabet Med
September 2025
Edinburgh Centre for Endocrinology & Diabetes, NHS Lothian, Edinburgh, UK.
Aims: This study aimed to assess the impact of the Omnipod 5 automated insulin delivery (AID) system on continuous glucose monitoring (CGM) metrics, HbA1c, and weight in a real-world setting. Additionally, independent predictors of glycaemic response were assessed.
Methods: Observational analysis of adults with type 1 diabetes using Omnipod 5 (n = 353).
G Ital Nefrol
August 2025
Infermiere Professionale SSD Nefrologia e Dialisi P.O. Soverato, ASP CZ.
Management of diabetes mellitus in hemodialysis is highly complex due to increased glycemic variability and hypoglycemic risk. The use of technologies applied to diabetes has been shown to improve glycemic control, however data in dialysis patients are limited. To describe the efficacy and safety of the minimed 780G AHCL system in a stable hemodialysis patient and during hospitalization in the Intensive Care Unit (ICU).
View Article and Find Full Text PDFProg Mol Biol Transl Sci
September 2025
Faculty of Dental Medicine and Oral Health Sciences, McGill University, Montreal, QC, Canada. Electronic address:
This chapter, "Implantable Biosensors: Advancements and Applications," provides a succinct overview of the state-of-the-art in implantable biosensor technology, highlighting both established clinical uses and promising areas of ongoing research. It begins by outlining the fundamental principles and advantages of these sensors, such as their precision in physiological monitoring and capability for real-time therapeutic interventions. A variety of implantable sensors are categorized, including biophysical and biochemical types, each designed for specific medical applications.
View Article and Find Full Text PDFDiabetes Res Clin Pract
September 2025
Health Education Administration, and Department of Endocrinology and Diabetes, Diabetes Treatment Center, Prince Sultan Military Medical City, Riyadh, Saudi Arabia.
Background: Despite advances, glycemic control in people with type 2 diabetes (PwT2D) treated with oral antidiabetic medications (ADMs) often remains suboptimal. Continuous glucose monitoring (CGM) has shown promise in diabetes management, offering real-time insights into glucose trends. This study evaluates the impact of transitioning from conventional self-monitoring of blood glucose (SMBG) to CGM on glycemic outcomes and self-management in PwT2D receiving oral ADMs.
View Article and Find Full Text PDF