Developing a 'multiPatchPET' system in GATE for a PET system design with irregular geometries.

Phys Med Biol

School of Biomedical Engineering, Shanghai Jiao Tong University, People's Republic of China. Department of Nuclear Medicine, Ruijin Hospital, Shanghai, People's Republic of China.

Published: September 2018


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

This work modified the commonly used Monte Carlo tool package GATE by developing a new 'multiPatchPET' system so that GATE users can easily simulate PET systems with irregular geometries. The motivation was to design a brain PET scanner with high sensitivity. It is known that compact PET scanners with a large solid coverage angle can achieve high sensitivity with fewer scintillation detectors, and thus have the potential to provide better image quality in brain PET imaging than conventional ring PET scanners. However, considering a straightforward example with the largest possible solid angle, a spherical PET scanner is hard to manufacture. A more practical alternative would be a sphere-like polyhedral PET scanner with flat detector patches. Moreover, when monolithic scintillators are chosen to construct these flat detector modules, detection efficiency is further improved. Thus, we plan to design a sphere-like polyhedral PET scanner made up of monolithic scintillators. Unfortunately, in our design study, we found that simulating such a scanner with the latest GATE version (8.0) was not trivial, since no predefined systems could be used. In this work we introduced a 'multiPatchPET' system to GATE, which we and other GATE users will be able to use to develop PET scanners with any irregular geometry and any shape of patch. To validate our modification, a single block detector and an mCT scanner were simulated via both the original 'ecat' system and the new 'multiPatchPET' system. The results show no difference in terms of the detecting efficiency and reconstruction image. Then we used the 'multiPatchPET' system to simulate an 86 surface polyhedral brain PET scanner. Compared with two cylindrical brain scanners, the polyhedral brain scanner shows a higher sensitivity and has fewer noisy images. Thus, it was proved that our modification, which is accessible to the nuclear imaging research community, equipped GATE with a powerful and user-friendly tool to simulate complex scanners with irregular patches easily.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aad8fdDOI Listing

Publication Analysis

Top Keywords

'multipatchpet' system
20
pet scanner
20
system gate
12
brain pet
12
pet scanners
12
pet
11
developing 'multipatchpet'
8
irregular geometries
8
gate users
8
scanner
8

Similar Publications

Developing a 'multiPatchPET' system in GATE for a PET system design with irregular geometries.

Phys Med Biol

September 2018

School of Biomedical Engineering, Shanghai Jiao Tong University, People's Republic of China. Department of Nuclear Medicine, Ruijin Hospital, Shanghai, People's Republic of China.

This work modified the commonly used Monte Carlo tool package GATE by developing a new 'multiPatchPET' system so that GATE users can easily simulate PET systems with irregular geometries. The motivation was to design a brain PET scanner with high sensitivity. It is known that compact PET scanners with a large solid coverage angle can achieve high sensitivity with fewer scintillation detectors, and thus have the potential to provide better image quality in brain PET imaging than conventional ring PET scanners.

View Article and Find Full Text PDF