A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Concordance between point-of-care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background: We tested the hypothesis that the results of the same test performed on point-of-care blood gas analysis (BGA) machine and automatic analyzer (AA) machine in central laboratory have high degree of concordance in critical care patients and that the two test methods could be used interchangeably.

Methods: We analyzed 9398 matched pairs of BGA and AA results, obtained from 1765 patients. Concentration pairs of the following analytes were assessed: hemoglobin, glucose, sodium, potassium, chloride, and bicarbonate. We determined the agreement using concordance correlation coefficient (CCC) and Bland-Altman analysis. The difference in results was also assessed against the United States Clinical Laboratory Improvement Amendments (US-CLIA) 88 rules. The test results were considered to be interchangeable if they were within the US-CLIA variability criteria and would not alter the clinical management when compared to each other.

Results: The median time interval between sampling for BGA and AA in each result pair was 5 minutes. The CCC values ranged from 0.89(95% CI 0.89-0.90) for chloride to 0.98(95% CI 0.98-0.99) for hemoglobin. The largest bias was for hemoglobin. The limits of agreement relative to bias were largest for sodium, with 3.4% of readings outside the US-CLIA variation rule. The number of readings outside the US-CLIA acceptable variation was highest for glucose (7.1%) followed by hemoglobin (5.9%) and chloride (5.2%).

Conclusion: We conclude that there is moderate to substantial concordance between AA and BGA machines on tests performed in critically ill patients. However, the two tests methods cannot be used interchangeably, except for potassium.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6817263PMC
http://dx.doi.org/10.1002/jcla.22425DOI Listing

Publication Analysis

Top Keywords

point-of-care blood
8
blood gas
8
gas analysis
8
critically ill
8
ill patients
8
readings us-clia
8
hemoglobin
5
concordance
4
concordance point-of-care
4
analysis laboratory
4

Similar Publications