98%
921
2 minutes
20
The use of an antimony film coated on a screen-printed carbon electrode (ex-situ SbSPCE) is proposed for the determination of Pd(II) at ppb levels in natural samples by adsorptive stripping voltammetry using dimethylglyoxime as chelating agent. Ex-situ SbSPCE produces a better analytical performance as compared to a commercially sputtered bismuth screen-printed electrode (BiSPE). The detection and quantification limits were 2.7 and 9.0µgL respectively with a good linear behaviour in the wide examined concentration range (from 1µgL up to 100.0µgL, R=0.998). The proposed ex-situ SbSPCE showed an excellent repeatability with a relative standard deviation of 0.5% for ten successive measurements and a very good reproducibility (1.6% for three different ex-situ SbSPCE units within series of ten repetitive assays). Moreover, the ex-situ SbSPCE was successfully applied for the determination of low concentration levels of Pd(II) in spiked tap water with a very high reproducibility (0.2%) and providing equivalent results to those achieved by ICP-MS measurements.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.talanta.2017.01.084 | DOI Listing |
Talanta
May 2017
Departament d'Enginyeria Química i Química Analítica, Facultat de Química, Universitat de Barcelona, Martí i Franquès 1-11, E-08028 Barcelona, Spain.
The use of an antimony film coated on a screen-printed carbon electrode (ex-situ SbSPCE) is proposed for the determination of Pd(II) at ppb levels in natural samples by adsorptive stripping voltammetry using dimethylglyoxime as chelating agent. Ex-situ SbSPCE produces a better analytical performance as compared to a commercially sputtered bismuth screen-printed electrode (BiSPE). The detection and quantification limits were 2.
View Article and Find Full Text PDF