Background: Quantification of glycated hemoglobin (HbA1c) is a challenge in patients with hemoglobin (Hb) variants. We evaluated the impact of various Hb variants on five routine HbA1c assays by comparing with the IFCC reference measurement procedure (RMP).
Methods: Whole blood samples showing warning flags or no results on routine HPLC HbA1c assays were confirmed for Hb variants and were submitted to HbA1c quantification using Sebia Capillarys 2 Flex Piercing, Roche Tina-quant HbA1c Gen. 2, Bio-Rad Variant II Turbo 2.0, ADAMS HA-8180, Tosoh G8 standard mode, and IFCC RMP using LC-MS.
Results: Among 114 samples, the most common variants were Hb G-Coushatta (n=47), Queens (n=41), Ube-4 (n=11), Chad (n=4), Yamagata (n=4), G-His-Tsou (n=2), G-Taipei (n=1), Fort de France (n=1), Hoshida (n=1), and two novel variants (Hb α-globin, HBA 52 Gly>Cys and Hb β-globin, HBB 146 His>Asn). In terms of control samples, all the result of HbA1c were "acceptable", within the criteria of ±7% compared to IFCC RMP target values. However, percentage of "unacceptable" results of samples with Hb variants were 16% for Capillarys 2, 7% for Tina-quant, 51% for Variant II Turbo 2.0, 95% for G8 standard mode, and 89% for HA-8180. The Capillarys 2 and HA-8180 assay did not provide the results in 5 and 40 samples with Hb variants, respectively.
Conclusions: HbA1c results from five routine assays in patients with relatively common Hb variants in Korea showed various degrees of bias compared to those of IFCC RMP. Therefore, laboratories should be aware of the limitation of their methods with respect to interference from Hb variants found commonly in their local population and suggest an alternative HbA1c quantification method.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/cclm-2016-0865 | DOI Listing |
Clin Chem Lab Med
September 2025
Population Health and Genomics, Ninewells Hospital and Medical School, University of Dundee, Dundee, UK.
Objectives: Accurate 25-hydroxyvitamin D (25-(OH)D) assays are essential for defining vitamin D status and ensuring appropriate clinical decisions. Standardization efforts, including the Vitamin D Standardization Program (VDSP), aim to minimize assay variability. This study evaluates the measurement uncertainty (MU) of various 25-(OH)D assays and their ability to detect physiologically relevant changes over time.
View Article and Find Full Text PDFJ Clin Lipidol
June 2025
Division of Laboratory Sciences, Centers for Disease Control and Prevention (CDC), Atlanta, GA, USA (Drs Lyle, Flores, Coffman, Doty, Sugahara, Vesper).
Background: Lipoprotein(a) [Lp(a)] is an independent risk factor for cardiovascular diseases (CVD). Recent clinical guidelines recommend measuring Lp(a); however, the lack of Lp(a) assay standardization presents challenges to using common clinical decision points. Assay standardization may minimize interassay variability.
View Article and Find Full Text PDFClin Chem Lab Med
February 2025
Roche Diagnostics GmbH, Penzberg, Germany.
Objectives: Accurate measurement of serum cortisol is crucial for the diagnosis and management of adrenal disorders. Thus, we have developed a novel isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC MS/MS)-based candidate reference measurement procedure (RMP) to quantify cortisol in human serum/plasma, offering higher sensitivity and reliability compared to existing RMPs.
Methods: Quantitative Nuclear Magnetic Resonance spectroscopic (qNMR) methodology has been utilized to assign the absolute content (g/g) and SI-traceability to the reference materials.
Anal Bioanal Chem
May 2025
Chemical Sciences Division, National Institute of Standards and Technology (NIST), Gaithersburg, MD, 20899, USA.
Commutability is where the measurement response for a reference material (RM) is the same as for an individual patient sample with the same concentration of analyte measured using two or more measurement systems. Assessment of commutability is essential when the RM is used in a calibration hierarchy or to ensure that clinical measurements are comparable across different measurement procedures and at different times. The commutability of three new Standard Reference Materials (SRMs) for determining serum total 25-hydroxyvitamin D [25(OH)D], defined as the sum of 25-hydroxyvitamin D [25(OH)D] and 25-hydroxyvitamin D [25(OH)D], was assessed through an interlaboratory study.
View Article and Find Full Text PDFJ Atheroscler Thromb
May 2025
Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center.
Aim: Lipoprotein (a) [Lp(a)] is a well-established risk factor for cardiovascular disease independent of low-density lipoprotein-cholesterol (LDL-C). The Lp(a) concentrations were inconsistent between the immunoassays. This study aimed to investigate whether harmonization of Lp(a) measurements can be achieved using a serum panel value assigned with the IFCC-endorsed mass spectrometry-based reference measurement procedure (IFCC-MS-RMP).
View Article and Find Full Text PDF