A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

Identifying activity levels and steps of people with stroke using a novel shoe-based sensor. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Background/purpose: Advances in sensor technologies provide a method to accurately assess activity levels of people with stroke in their community. This information could be used to determine the effectiveness of rehabilitation interventions as well as provide behavior-enhancing feedback. The purpose of this study was to assess the accuracy of a novel shoe-based sensor system (SmartShoe) to identify different functional postures and steps in people with stroke. The SmartShoe system consists of five force-sensitive resistors built into a flexible insole and an accelerometer on the back of the shoe. Pressure and acceleration data are sent via Bluetooth to a smart phone.

Methods: Participants with stroke wore the SmartShoe while they performed activities of daily living (ADLs) in sitting, standing, and walking positions. Data from four participants were used to develop a multilayer perceptron artificial neural network (ANN) to identify sitting, standing, and walking. A signal-processing algorithm used data from the pressure sensors to estimate the number of steps taken while walking. The accuracy, precision, and recall of the ANN for identifying the three functional postures were calculated with data from a different set of participants. Agreement between steps identified by SmartShoe and actual steps taken was analyzed by the Bland Altman method.

Results: The SmartShoe was able to accurately identify sitting, standing, and walking. Accuracy, precision, and recall were all greater than 95%. The mean difference between steps identified by SmartShoe and actual steps was less than one step.

Discussion: The SmartShoe was able to accurately identify different functional postures, using a unique combination of pressure and acceleration data, of people with stroke as they performed different ADLs. There was a strong level of agreement between actual steps taken and steps identified by the SmartShoe. Further study is needed to determine whether the SmartShoe could be used to provide valid information on activity levels of people with stroke while they go about their daily lives in their home and community.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3355328PMC
http://dx.doi.org/10.1097/NPT.0b013e318256370cDOI Listing

Publication Analysis

Top Keywords

people stroke
20
activity levels
12
functional postures
12
sitting standing
12
standing walking
12
steps identified
12
identified smartshoe
12
actual steps
12
steps
9
smartshoe
9

Similar Publications