A real-time respiratory motion monitoring system using KINECT: proof of concept.

Med Phys

Department of Radiation Oncology, University of Iowa, Iowa City, IA 52242, USA.

Published: May 2012


Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Purpose: The purpose of this study is to investigate the feasibility of a low-cost respiratory motion monitoring system based on the Microsoft KINECT sensor.

Methods: The authors increased KINECT's inherent depth resolution from 1 cm to 1 mm via a motion magnification system. Using the KINECT software development kit, the authors programmed the KINECT to capture depth images and determine the average depth over a thoracic region of interest, viewed almost parallel to the subject's surface. KINECT respiratory traces (average depth vs time at a rate of 30 Hz) were acquired from four volunteers and compared with those simultaneously acquired using a commercially available strain gauge respiratory gating system.

Results: The correlation coefficient (CC) between KINECT and strain gauge traces varied from 0.958 to 0.978, with a mean CC of 0.969. This strong correlation was also demonstrated by the joint probability distribution and visual inspection.

Conclusions: It is feasible to use the KINECT for respiratory motion tracking. Traces are similar to those of a clinically used strain gauge system. The KINECT-based system provides a new and economical way to monitor respiratory motion.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4704644DOI Listing

Publication Analysis

Top Keywords

respiratory motion
16
strain gauge
12
motion monitoring
8
monitoring system
8
system kinect
8
average depth
8
kinect respiratory
8
kinect
7
motion
5
system
5

Similar Publications

Background: In clinical practice, digital subtraction angiography (DSA) often suffers from misregistration artifact resulting from voluntary, respiratory, and cardiac motion during acquisition. Most prior efforts to register the background DSA mask to subsequent postcontrast images rely on key point registration using iterative optimization, which has limited real-time application.

Purpose: Leveraging state-of-the-art, unsupervised deep learning, we aim to develop a fast, deformable registration model to substantially reduce DSA misregistration in craniocervical angiography without compromising spatial resolution or introducing new artifacts.

View Article and Find Full Text PDF

Background: Understanding respiratory motions of liver and its surrogate organs is crucial for precise dose delivery in liver cancer radiotherapy. Although these motions have been studied for respiratory motion management in the supine posture, few studies have quantified them and evaluated their correlations in the upright posture.

Purpose: This study quantified the respiratory motions of liver and surrogate organs and evaluated the correlations between the liver motions and surrogate signals for respiratory motion monitoring in both the supine and upright postures.

View Article and Find Full Text PDF

Background: Four-dimensional magnetic resonance imaging (4D-MRI) holds great promise for precise abdominal radiotherapy guidance. However, current 4D-MRI methods are limited by an inherent trade-off between spatial and temporal resolutions, resulting in compromised image quality characterized by low spatial resolution and significant motion artifacts, hindering clinical implementation. Despite recent advancements, existing methods inadequately exploit redundant frame information and struggle to restore structural details from highly undersampled acquisitions.

View Article and Find Full Text PDF

Understanding gastric physiology in rodents is critical for advancing preclinical neurogastroenterology research. However, existing techniques are often invasive, terminal, or limited in resolution. This study aims to develop a non-invasive, standardized MRI protocol capable of capturing whole-stomach dynamics in anesthetized rats with high spatiotemporal resolution.

View Article and Find Full Text PDF

Background: Stereotactic body radiotherapy (SBRT) is an effective treatment for early-stage non-small cell lung cancer. However, patient breathing can affect treatment accuracy. Therefore, this study aimed to develop a bi-polar (BP) gated motion management strategy for SBRT and evaluate its feasibility geometrically and dosimetrically.

View Article and Find Full Text PDF