Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objective: The Zenith Renu abdominal aortic aneurysm (AAA) Ancillary Graft (Cook Medical Inc, Bloomington, Ind) provides active proximal fixation for treatment of pre-existing endografts with failed or failing proximal fixation or seal. The purpose of this study was to evaluate the midterm outcomes of treatment with this device.

Methods: From September 2005 to November 2006, a prospective, nonrandomized, multicenter, postmarket registry was utilized to collect physician experiences from 151 cases (89 converters and 62 main body extensions) at 95 institutions. Preoperative indications and procedural and postimplantation outcomes were collected and analyzed. Technical success and clinical success were determined as defined by the Society of Vascular Surgery reporting standards.

Results: Patients were predominantly male (87%) with a mean age of 77 years. The interval between the original endograft implantation to Renu treatment was 43.4 ± 18.7 months. The indications for treatment were endoleak (n = 111), migration (n = 136), or both (n = 94). Technical success was 98.0% with two cases of intraoperative conversion and one case of persistent type IA endoleak. The median follow-up for the cohort was 45.0 months (range, 0-56 months; interquartile range, 25.0 months). Overall, 32 cases had treatment failures that included at least one of the following: death (n = 5), type I/III endoleak (n = 18), graft infection (n = 1), thrombosis (n = 1), aneurysm enlargement >5 mm (n = 9), rupture (n = 4), conversion (n = 9, with 7 after 30 days), and migration (n = 1). Overall, the clinical success for the entire cohort during the follow-up period was 78.8% (119/151).

Conclusions: The postmarket registry data confirm that the Zenith Renu AAA Ancillary Graft can be used to treat endovascular repairs that failed due to proximal attachment failures. The salvage treatment with the Renu device had high technical success rate and resulted in clinical success in a majority of patients (78.8%). While failed endovascular repairs can be salvaged, a clinical failure in one of five patients still emphasizes the importance of patient and device selection during initial endovascular aneurysm repair to ensure durable success.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2010.12.058DOI Listing

Publication Analysis

Top Keywords

zenith renu
12
aaa ancillary
12
ancillary graft
12
technical success
12
clinical success
12
midterm outcomes
8
renu aaa
8
proximal fixation
8
postmarket registry
8
endovascular repairs
8

Similar Publications

Objective: The physical impact of the application of Heli-FX EndoAnchors (EA; Medtronic, Minneapolis, USA) on endograft (EG) material is unclear. This study aimed to examine the possible EG membrane alterations after EA implantation.

Methods: Heli-FX EndoAnchors were applied into four aortic endocuffs: AFX2 (Endologix Inc.

View Article and Find Full Text PDF

Objective: Secondary interventions after endovascular aneurysm repair (EVAR) remain a concern. Most are simple catheter-based procedures, but in some instances, open conversions (OCs) are required and carry a worse outcome. We reviewed our experience to characterize these OCs.

View Article and Find Full Text PDF

Objective: The Zenith Renu abdominal aortic aneurysm (AAA) Ancillary Graft (Cook Medical Inc, Bloomington, Ind) provides active proximal fixation for treatment of pre-existing endografts with failed or failing proximal fixation or seal. The purpose of this study was to evaluate the midterm outcomes of treatment with this device.

Methods: From September 2005 to November 2006, a prospective, nonrandomized, multicenter, postmarket registry was utilized to collect physician experiences from 151 cases (89 converters and 62 main body extensions) at 95 institutions.

View Article and Find Full Text PDF

Background: The explosion in endovascular interventions for peripheral vascular disease has resulted in procedures being used by a multitude of specialties. Nonvascular surgeons performing these interventions can create scenarios that may make future vascular interventions difficult. In this article, we present a case report illustrating this point.

View Article and Find Full Text PDF

A comparative analysis of the outcomes of aortic cuffs and converters for endovascular graft migration.

J Vasc Surg

June 2010

Section of Vascular Surgery, Division of General Surgery, Department of Surgery, Washington University School of Medicine, Barnes-Jewish Hospital, St. Louis, Mo, USA.

Objective: Proximal attachment failure, often leading to graft migration, is a severe complication of endovascular aneurysm repair (EVAR). Aortic cuffs have been used to treat proximal attachment failure with mixed results. The Zenith Renu AAA Ancillary Graft (Cook Inc, Bloomington, Ind) is available in two configurations: converter and main body extension.

View Article and Find Full Text PDF