Objectives: We evaluated the efficiency and safety of the Occlutech Figulla device in percutaneous closure of secundum atrial septal defects (ASD).

Study Design: The study included 28 patients (17 women, 11 men; mean age 43 years) who underwent percutaneous transcatheter closure using the Occlutech Figulla device for secundum ASDs causing a hemodynamically significant shunt. Defect size was estimated by transthoracic (TTE) and transesophageal (TEE) echocardiography, and also by balloon sizing in nine patients. The patients were followed-up for six months and were examined by TTE.

Results: The mean defect size was 20.3+/-2.1 mm on TTE, 22.1+/-1.9 mm on TEE, and 24.2+/-2.4 mm on balloon sizing. The mean device size was 26.8+/-3.6 mm (range 6 to 36 mm). The mean procedure time was 44.7+/-21.4 minutes. The device was placed successfully in all the patients. A small residual flow was seen immediately after device placement in three patients (10.7%), which disappeared in two patients at three months, and in one patient at six months. During the procedure, complications were seen in four patients (14.3%), including transient sinus tachycardia in three patients (10.7%) and acute atrial fibrillation in one patient (3.6%). At six months, all the patients were asymptomatic. No ischemic stroke, cardiac perforation, device erosion, embolization, thrombus formation, or malposition of the device were observed.

Conclusion: The Occlutech Figulla occluder is a safe and efficient device to close secundum ASDs. It may be preferred especially in patients with a high risk for thrombus formation.

Download full-text PDF

Source

Publication Analysis

Top Keywords

occlutech figulla
16
figulla device
12
patients
10
device
9
efficiency safety
8
percutaneous closure
8
closure secundum
8
secundum atrial
8
atrial septal
8
septal defects
8

Similar Publications

Background: This study investigated current trends in transcatheter atrial septal defect (ASD) closure among adult patients, with an emphasis on device and size selection, as well as acute complications.

Methods And Results: This study used the Japanese Structural Heart Disease (J-SHD) registry database, which is a prospective nationwide multicenter registry. In all, 1,921 patients who underwent transcatheter ASD closure between 2019 and 2022 were analyzed in this study.

View Article and Find Full Text PDF

Background: Transcatheter atrial septal defect (ASD) closure is the first treatment option for secundum ASD, but parameters for optimal device selection have not been established. We compared outcomes between occluders with a wire frame and metal mesh devices.

Methods And Results: This study included secundum ASD patients implanted with a wire frame occluder (GORECARDIOFORM ASD occluder [GCA]; W.

View Article and Find Full Text PDF

Percutaneous closure of atrial septal defect (ASD) is less invasive than surgical closure and yields good results; however, cardiac erosion is a serious complication. It usually occurs within 72 h after implantation and rarely after months to years. Cardiac erosion has been rarely reported using Figulla Flex II (FFII; Occlutech, Schaffhausen, Switzerland).

View Article and Find Full Text PDF

We describe a unique report of percutaneous closure of multiple secundum atrial septal defects in a child utilizing three Occlutech Figulla septal occluders deployed sequentially. The procedure was performed under live three-dimensional transesophageal echocardiography guidance.

View Article and Find Full Text PDF

Transcatheter closure of a patent foramen ovale (PFO) is performed in cryptogenic stroke and other conditions. Information is lacking for some devices. We aimed to evaluate the Figulla Flex II PFO Occluder (FFP) and Figulla Flex UNI Occluder (FFU) through a retrospective multi-center registry.

View Article and Find Full Text PDF