Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 197
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 317
Function: require_once
98%
921
2 minutes
20
Objective: To assess the impedance cardiogram recorded by an automated external defibrillator during cardiac arrest to facilitate emergency care by lay persons. Lay persons are poor at emergency pulse checks (sensitivity 84%, specificity 36%); guidelines recommend they should not be performed. The impedance cardiogram (dZ/dt) is used to indicate stroke volume. Can an impedance cardiogram algorithm in a defibrillator determine rapidly circulatory arrest and facilitate prompt initiation of external cardiac massage?
Design: Clinical study.
Setting: University hospital.
Patients: Phase 1 patients attended for myocardial perfusion imaging. Phase 2 patients were recruited during cardiac arrest. This group included nonarrest controls.
Interventions: The impedance cardiogram was recorded through defibrillator/electrocardiographic pads oriented in the standard cardiac arrest position.
Measurements And Main Results: Phase 1: Stroke volumes from gated myocardial perfusion imaging scans were correlated with parameters from the impedance cardiogram system (dZ/dt(max) and the peak amplitude of the Fast Fourier Transform of dZ/dt between 1.5 Hz and 4.5 Hz). Multivariate analysis was performed to fit stroke volumes from gated myocardial perfusion imaging scans with linear and quadratic terms for dZ/dt(max) and the Fast Fourier Transform to identify significant parameters for incorporation into a cardiac arrest diagnostic algorithm. The square of the peak amplitude of the Fast Fourier Transform of dZ/dt was the best predictor of reduction in stroke volumes from gated myocardial perfusion imaging scans (range = 33-85 mL; p = .016). Having established that the two pad impedance cardiogram system could detect differences in stroke volumes from gated myocardial perfusion imaging scans, we assessed its performance in diagnosing cardiac arrest. Phase 2: The impedance cardiogram was recorded in 132 "cardiac arrest" patients (53 training, 79 validation) and 97 controls (47 training, 50 validation): the diagnostic algorithm indicated cardiac arrest with sensitivities and specificities (+/- exact 95% confidence intervals) of 89.1% (85.4-92.1) and 99.6% (99.4-99.7; training) and 81.1% (77.6-84.3) and 97% (96.7-97.4; validation).
Conclusions: The impedance cardiogram algorithm is a significant marker of circulatory collapse. Automated defibrillators with an integrated impedance cardiogram could improve emergency care by lay persons, enabling rapid and appropriate initiation of external cardiac massage.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/CCM.0b013e3181c02ca1 | DOI Listing |