A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 197

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 197
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 271
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3165
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 597
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 511
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 317
Function: require_once

The impedance cardiogram recorded through two electrocardiogram/defibrillator pads as a determinant of cardiac arrest during experimental studies. | LitMetric

Category Ranking

98%

Total Visits

921

Avg Visit Duration

2 minutes

Citations

20

Article Abstract

Objective: Laypersons are poor at emergency pulse checks (sensitivity 84%, specificity 36%). Guidelines indicate that pulse checks should not be performed. The impedance cardiogram (dZ/dt) is used to assess stroke volume. Can a novel defibrillator-based impedance cardiogram system be used to distinguish between circulatory arrest and other collapse states?

Design: Animal study.

Setting: University research laboratory.

Subjects: Twenty anesthetized, mechanically ventilated pigs, weight 50-55 kg.

Interventions: Stroke volume was altered by right ventricular pacing (160, 210, 260, and 305 beats/min). Cardiac arrest states were then induced: ventricular fibrillation (by rapid ventricular pacing) and, after successful defibrillation, pulseless electrical activity and asystole (by high-dose intravenous pentobarbitone).

Measurements And Main Results: The impedance cardiogram was recorded through electrocardiogram/defibrillator pads in standard cardiac arrest positions. Simultaneously recorded electro- and impedance cardiogram (dZ/dt) along with arterial blood pressure tracings were digitized during each pacing and cardiac arrest protocol. Five-second epochs were analyzed for sinus rhythm (20 before ventricular fibrillation, 20 after successful defibrillation), ventricular fibrillation (40), pulseless electrical activity (20), and asystole (20), in two sets of ten pigs (ten training, ten validation). Standard impedance cardiogram variables were noncontributory in cardiac arrest, so the fast Fourier transform of dZ/dt was assessed. During ventricular pacing, the peak amplitude of fast Fourier transform of dZ/dt (between 1.5 and 4.5 Hz) correlated with stroke volume (r2 = .3, p < .001). In cardiac arrest, a peak amplitude of fast Fourier transform of dZ/dt of < or = 4 dB x ohm x rms indicated no output with high sensitivity (94% training set, 86% validation set) and specificity (98% training set, 90% validation set).

Conclusions: As a powerful clinical marker of circulatory collapse, the fast Fourier transformation of dZ/dt (impedance cardiogram) has the potential to improve emergency care by laypersons using automated defibrillators.

Download full-text PDF

Source
http://dx.doi.org/10.1097/CCM.0b013e318170a03bDOI Listing

Publication Analysis

Top Keywords

impedance cardiogram
28
cardiac arrest
24
fast fourier
16
stroke volume
12
ventricular pacing
12
ventricular fibrillation
12
fourier transform
12
transform dz/dt
12
cardiogram recorded
8
recorded electrocardiogram/defibrillator
8

Similar Publications