Purpose: To evaluate the retention of yellow Hader clips after exposure to various denture cleansers.

Methods: Seven groups of 18 yellow Hader clips each were soaked for the equivalent of 6 months of clinical use in the following denture cleansing solutions: Polident Regular, Polident Overnight, Efferdent, 5.25% Sodium Hypochlorite (NaOCl, 1:10 dilution) 15 min/day, NaOCl (1:10 dilution) 8 hours/day, water and dry (control) group. A Universal Testing Machine, set at a crosshead speed of 2 in/min, pulled each clip once, and the peak load-to-dislodgement was recorded and used as a measure to reflect changes in the retention of the Hader clips. Data were analyzed by a one-way analysis of variance followed by Tukey's HSD test. A p value < or = 0.05 was considered significant.

Results: Denture cleansing solutions affected the retentive values of yellow Hader clips (F= 6.102, p< or = .0001). Sodium hypochlorite solution, 15 min/day for 6 months, caused an increase in the retentive values of the clips tested with a mean peak load-to-dislodgement of 22.63 +/- 1.29 N. In addition, clips soaked in water showed no difference in retentive values when compared with all other groups. Furthermore, Polident Regular, Polident Overnight, Efferdent, and NaOCl (8 hours/day) had no effect on the retentive values of yellow Hader clips.

Conclusion: This in vitro study demonstrated that the retention of yellow Hader clips used in implant overdentures is unaffected when soaked in commercial effervescent denture cleansers (Polident 5 Minute, Polident Overnight, and Efferdent) for six simulated months. Sodium hypochlorite statistically increased the single-pull retentive values of the clips, an effect that may not be beneficial. Increased retentive values may be associated with reduced durability of clips; further research is needed to address this issue.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1532-849X.2006.00167.xDOI Listing

Publication Analysis

Top Keywords

yellow hader
24
hader clips
24
retentive values
24
denture cleansing
12
cleansing solutions
12
retention yellow
12
polident overnight
12
overnight efferdent
12
sodium hypochlorite
12
clips
10

Similar Publications

Purpose: To compare the retention forces of implant overdenture patrices (ball, bar, and TiSi.snap) to conventional (O-ring, metal housing, and clip) and polyvinyl siloxane (PVS)-based silicone (retention.sil 200, 400, and 600; Mucopren Soft; and GC Reline Soft) matrix materials.

View Article and Find Full Text PDF

The influence of saliva pH value on the retention and durability of bar-clip attachments.

J Adv Prosthodont

February 2015

Department of Rehabilitation and Maxillofacial Prostheses, Faculty of Odontology, University of Barcelona, Barcelona, Spain.

Purpose: The aim of this study was to compare the durability and retention of 4 types of attachments placed over computer aided design/computer aided manufacturing (CAD/CAM) titanium bars when subjected to different pH conditions.

Materials And Methods: Four commercially available attachments were investigated: Hader Yellow, Hader Red, Ackerman Gold and Ackerman Stainless Steel. These attachments and Ackerman CAD/CAM titanium bars were placed in 2 vessels containing different artificial saliva solutions (pH 7/pH 4) at 37℃ for one month to simulate corrosion conditions, and they were then subjected to mechanical testing (5400 cycles of insertion and removal).

View Article and Find Full Text PDF

Purpose: The aim of this study was to evaluate the effects of different implant angulations and attachment configurations on the retentive forces of locator attachment-retained overdentures over a simulated time period of 6 months.

Materials And Methods: Two acrylic resin models (models 1 and 2) were fabricated to represent an edentulous mandible. In model 1, the midline implant (in the central region) was vertical and the other two implants (in the canine regions) were 20 degrees divergent from the midline, reaching a total divergence of 40 degrees.

View Article and Find Full Text PDF

Purpose: The aim of this study was to examine early changes in retentive values of implant overdenture attachments during multiple pulls.

Materials And Methods: Two implant attachment systems (Hader bar and clip, Locator system) were used in this study. The experimental groups were divided into yellow Hader clips, white Locator attachments, and green Locator attachments.

View Article and Find Full Text PDF

This in vitro study evaluated the 6-month fatigue retention provided by 5 paired overdenture attachments placed at 3 different interimplant distances (19, 23, and 29 mm). Mean fatigue retention was calculated for each attachment type and compared with the retention produced by the other attachments tested and the initial retention values published earlier. Interimplant distance was found to play a significant role only in the retention of the Hader bar with red and yellow clips.

View Article and Find Full Text PDF