Background And Purpose: Next generation cone beam computed tomography (CBCT) technology has shown improved Hounsfield unit accuracy over standard CBCT, and comparable to that of fan beam CT simulators (CTsim), potentially allowing direct dose calculation on CBCT. In this study, we evaluated the dose calculation accuracy of clinical treatment plans calculated using this technology, compared with standard CBCT.

Materials And Methods: Thirty subjects with thoracic or upper abdominal cancer were imaged on CTsim, standard CBCT and next generation CBCT, and treated using breath hold techniques.

View Article and Find Full Text PDF

Purpose: Cone beam computed tomography (CBCT)-based online adaptive radiation therapy is carried out using a synthetic CT (sCT) created through deformable registration between the patient-specific fan-beam CT, fan-beam computed tomography (FBCT), and daily CBCT. Ethos 2.0 allows for plan calculation directly on HyperSight CBCT and uses artificial intelligence-informed tools for daily contouring without the use of a priori information.

View Article and Find Full Text PDF

Introduction: This work presents a multi-institutional study on image quality provided by a novel cone beam computed tomography (CBCT). The main goal is to investigate the consistency of imaging performance across multiple institutions.

Methods: Phantoms for measuring relative electron density (RED) and image quality were sent to six institutions for imaging on Ethos and Halcyon units equipped with HyperSight CBCT.

View Article and Find Full Text PDF

Background: Stereotactic arrythmia radioablation (STAR) is a noninvasive technique to treat ventricular tachycardia (VT). Management of cardiorespiratory motion plays an essential role in VT-STAR treatments to improve treatment outcomes by reducing positional uncertainties and increasing dose conformality. Use of an electrocardiogram (ECG) signal, acquired in real-time, as a surrogate to gate the beam has the potential to fulfil that intent.

View Article and Find Full Text PDF

Purpose: The presence of metal implants can produce artifacts and distort Hounsfield units (HU) in patient computed tomography (CT) images. The purpose of this work was to characterize a novel metal artifact reduction (MAR) algorithm for reconstruction of CBCT images obtained by the HyperSight imaging system.

Methods: Three tissue-equivalent phantoms were fitted with materials commonly used in medical applications.

View Article and Find Full Text PDF

. This work investigates the small-field dosimetric characteristics of a 2.5 MV sintered diamond target beam and its feasibility for use in linac-based intracranial stereotactic treatments.

View Article and Find Full Text PDF

The expectation of quality and safety is a fundamental tenet in all areas of healthcare, and a cornerstone of best practice is a process of continuous learning and continuous improvement. Independent audits and peer review of radiotherapy programs are an important mechanism for identifying process or technology gaps, for highlighting areas for improvement, and for incorporating within continuous improvement processes. In the field of radiotherapy, independent certification programs exist within various national and/or professional spheres, yet few focus specifically on specialty procedures such as radiosurgery or brachytherapy, despite several recommendations for such programs.

View Article and Find Full Text PDF

Background: Novel on-board CBCT allows for improved image quality and Hounsfield unit accuracy. When coupled with online adaptive tools, this may have potential to allow for simulation and treatment to be completed in a single on-table session.

Purpose: To study the feasibility of a high-efficiency radiotherapy treatment workflow without the use of a separate session for simulation imaging.

View Article and Find Full Text PDF

Purpose: The goal of this study was to evaluate the image quality provided by a novel cone beam computed tomography (CBCT) platform (HyperSight, Varian Medical Systems), a platform with enhanced reconstruction algorithms as well as rapid acquisition times. Image quality was compared with both status quo CBCT for image guidance, and to fan beam CT (FBCT) acquired on a CT simulator (CTsim).

Methods And Materials: In a clinical study, 30 individuals were recruited for whom either deep inspiration (DIBH) or deep exhalation breath hold (DEBH) was used during imaging and radiation treatment of tumors involving liver, lung, breast, abdomen, chest wall, and pancreatic sites.

View Article and Find Full Text PDF

Purpose: Approximately 90% of patients undergoing breast cancer radiation therapy experience skin toxicities that are difficult to classify and predict ahead of time. A prediction of toxicity at the early stages of the treatment would provide clinicians with a prompt to intervene. The objectives of this study were to evaluate the correlation between skin toxicity and radiomic features extracted from optical and infrared (thermal) images of skin, and to develop a model for predicting a patient's skin response to radiation.

View Article and Find Full Text PDF

Purpose And Objective: To describe a methodology for the dual-material fused deposition modeling (FDM) 3D printing of plastic scintillator arrays, to characterize their light output under irradiation using an sCMOS camera, and to establish a methodology for the dosimetric calibration of planar array geometries.

Materials And Methods: We have published an investigation into the fabrication and characterization of single element FDM printed scintillators intending to produce customizable dosimeters for radiation therapy applications. This work builds on previous investigations by extending the concept to the production of a high-resolution (scintillating element size 3 × 3 × 3 mm ) planar scintillator array.

View Article and Find Full Text PDF

Purpose: To investigate the feasibility of cardiac synchronized gating in stereotactic body radiation therapy (SBRT) of ventricular tachycardia (VT) using a real-time electrocardiogram (ECG) signal acquisition.

Methods And Materials: Stability of beam characteristics during simulated ECG gating was examined by developing a microcontroller interface to a Varian Clinac iX linear accelerator allowing gating at frequencies and duty cycles relevant to cardiac rhythm. Delivery accuracy was evaluated by measuring dose linearity with an ionization chamber, and flatness and symmetry with a two-dimensional detector array, for different gating windows within typical human cardiac cycle periods.

View Article and Find Full Text PDF

Entry into the field of clinical medical physics is most commonly accomplished through the completion of a Commission on Accreditation of Medical Physics Educational Programs (CAMPEP)-accredited graduate and residency program. To allow a mechanism to bring valuable expertise from other disciplines into clinical practice in medical physics, an "alternative pathway" approach was also established. To ensure those trainees who have completed a doctoral degree in physics or a related discipline have the appropriate background and didactic training in medical physics, certificate programs and a CAMPEP-accreditation process for these programs were initiated.

View Article and Find Full Text PDF

Purpose: Over the past decades, continuous efforts have been made to improve megavoltage (MV) image quality versus dose characteristics, including the implementation of low atomic number (Z) targets in MV beamlines and the development of more efficient detectors. Recently, a diamond target beam within a commercial radiotherapy treatment platform demonstrated improved planar contrast-to-noise-ratio (CNR) per unit dose using a novel 2.5 MV sintered diamond target beam, which enabled image acquisition on the order of mGy.

View Article and Find Full Text PDF

Purpose: Thermoplastic immobilizers are used routinely in radiation therapy to achieve positioning accuracy. These devices are variable in quality as they are dependent on the skill of the human fabricator. We examine the potential multi jet fusion (MJF) 3D printing for the production immobilizers with a focus on the surface dosimetry of several MJF-printed PA12-based material candidates.

View Article and Find Full Text PDF

Radiotherapy (RT) is an effective cancer treatment modality, but standard RT often causes collateral damage to nearby healthy tissues. To increase therapeutic ratio, radiosensitization via gold nanoparticles (GNPs) has been shown to be effective. One challenge is that megavoltage beams generated by clinical linear accelerators are poor initiators of the photoelectric effect.

View Article and Find Full Text PDF

Brachytherapy is a mature treatment modality. The literature is abundant in terms of review articles and comprehensive books on the latest established as well as evolving clinical practices. The intent of this article is to part ways and look beyond the current state-of-the-art and review emerging technologies that are noteworthy and perhaps may drive the future innovations in the field.

View Article and Find Full Text PDF

Purpose: To investigate the possible advantages of using 4pi-optimized arc trajectories in stereotactic body radiation therapy of ventricular tachycardia (VT-SBRT) to minimize exposure of healthy tissues.

Methods And Materials: Thorax computed tomography (CT) data for 15 patients were used for contouring organs at risk (OARs) and defining realistic planning target volumes (PTVs). A conventional trajectory plan, defined as two full coplanar arcs was compared to an optimized-trajectory plan provided by a 4pi algorithm that penalizes geometric overlap of PTV and OARs in the beam's-eye-view.

View Article and Find Full Text PDF

Purpose: The purpose of this study is to improve dose distribution and organ-at-risk sparing during gynecologic HDR brachytherapy with patient-specific applicators. The majority of applicators used today are generic in design and do not allow for dose modulation for patient-specific shaping of dose distributions. Their performance might be adjusted with commercially available wedge shields; however, this provides dose modulation in the orthogonal plane only and does not allow for variation along the length of the applicator.

View Article and Find Full Text PDF

Background And Purpose: A commercial 2.5 MV beam has been clinically available for beam's-eye-view imaging in radiotherapy, offering improved contrast-to-noise ratio (CNR) compared to therapeutic beams, due to the softer spectrum. Previous research suggested that imaging performance could be improved using a low-Z diamond target to reduce the self-absorption of diagnostic energy photons.

View Article and Find Full Text PDF

We propose a new methodology for the fabrication and evaluation of scintillating detector elements using a consumer grade fusion deposition modeling (FDM) 3D printer. In this study we performed a comprehensive investigation into both the effects of the 3D printing process on the scintillation light output of 3D printed plastic scintillation dosimeters (PSDs) and their associated dosimetric properties. Fabrication properties including print variability, layer thickness, anisotropy and extrusion temperature were assessed for 1 cm printed samples.

View Article and Find Full Text PDF

Unlabelled: This work introduces a novel capacitive-sensing technology capable of detecting respiratory motion with high temporal frequency (200 Hz). The system does not require contact with the patient and has the capacity to sense motion through clothing or plastic immobilization devices.

Abstract: PURPOSE: This work presents and evaluates a novel capacitive monitoring system (CMS) technology for continuous detection of respiratory motion during radiation therapy.

View Article and Find Full Text PDF

Purpose/objectives: Three-dimensional (3D) printing is recognized as an effective clinical and educational tool in procedurally intensive specialties. However, it has a nascent role in radiation oncology. The goal of this investigation is to clarify the extent to which 3D printing applications are currently being used in radiation oncology through a systematic review of the literature.

View Article and Find Full Text PDF

Brachytherapy accelerated partial breast irradiation (APBI) is well tolerated, but reported acute toxicities including moist desquamation rates range from 7% to 39%. Moist desquamation is correlated to long-term skin toxicity and high skin dose is the main risk factor. This study uses radiochromic films for in vivo skin dosimetry of low dose rate (LDR) APBI brachytherapy and prediction of skin toxicity.

View Article and Find Full Text PDF

The purpose of this investigation is to improve intra-fractional motion detection during cranial stereotactic radiosurgery with a novel capacitive motion sensing (CMS) system. Previous work showed that a capacitive detection system, based on a MPR121 capacitance-to-digital converter, provided a number of advantages over existing patient imaging systems used in the clinic, by uniquely offering ionizing-radiation-free and continuous monitoring without modification to the immobilization mask or treatment room. However, in order to provide submillimeter detection accuracy, the MPR121-based CMS system required relatively large sensors in close proximity to the patient.

View Article and Find Full Text PDF